期刊文献+

双能X射线骨密度仪测定技术研究 被引量:7

THE RESEARCH ON MEASURING PROCEDURE OF DUAL ENERGY X-RAY ABSORPTIOMETRY IN DETERMINING THE BONE MINERAL DENSITY
下载PDF
导出
摘要 以Lunar公司DPX-L型双能X射线骨密度仪为例,研究引起测量误差的主要因素并提出质控措施。用自检模块、腰椎体模评定仪器的体外测量精度好于1%。对10位志愿者重复2次测量,体内精度对腰椎,MPD(平均百分差)为0.46,CV变异系数)为0.75;对股骨颈,MPD为1.26、CV为0.47;Ward’s三角区MPD为2.01、CV为1.14;大转子MPD为1.66、CV为0.72。扫描图像由2位技术人员分析比由同1人分析时的误差大。测量对象体位重复和分析技术的规范化,是减小测量误差的2个重要的技术保障。为满足数据比对的需要,建议建立我国自己的体模标准。 The influence of measuring procedure on precision in determing the bone mineral density (BMD) using dual energy X-ray absorptionmetry (DXA) is studied. The research results show that the position of measured subjects and the analysis of acqured images are two main influence facfors. The precision of BMD values obtained from images analysed by two technicians is lower than that by one technician. For DXA of Lunar DPX-L, the vitro precision is better than 1 %. The vivo precision expressed by coefficient of variation (CV) are 0. 75% at lumbar, 0. 47 % at neck, 1. 14 % at Ward's triangle and 0. 72 % at trochanter, respectively. A suggestion for establishment of standard phantom of our country is proposed in the paper.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 1996年第3期230-235,共6页 Atomic Energy Science and Technology
关键词 骨密度 双能X线吸收术 体外精度 体内精度 Bone mineral density Dual energy X-ray absorptiometry Vitro precision Vivo precision
  • 相关文献

同被引文献58

引证文献7

二级引证文献77

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部