期刊文献+

β_2-微球蛋白试剂对尿素测定结果的影响 被引量:1

Interfering Effect of β_2-microglobulin Reagent on Urea Assay Results
下载PDF
导出
摘要 目的研究对尿素自动分析产生污染的项目及解决方法。方法将筛查出对尿素测定有明显影响的β2-微球蛋白试剂加入混合血清,测定其尿素,并与加入生理盐水的对照管进行比较,重复10次,观察分析β2-微球蛋白试剂对尿素测定结果的影响程度。结果β2-微球蛋白试剂对尿素对尿素测定结果产生明显的正干扰。结论自动生化分析中应注意测试项之间试剂对检测结果的影响,可用带模块的生化分析仪排除干扰,或者利用仪器的清洁程序,消除干扰。 Objective To investigate which regent that interfering on the urea auto- assay results, and to find the resolve measures of this effect. Methods β2- microglobulin reagent which was be screened to have apparent interfering effect on the urea assay results was be added into mixed serum. Urea concentration was detected for 10 times within the mixture and the normal control which was added with normal saline, and the effect of β2 - microglobulin reagent to urea results was assessed. Conclusion The influencing effect of β2 - microglobulin reagent on the urea assay results is significant and must be considered in auto clinical chemistry assay. It can be eliminated by clinical chemistry analyzer with modular or the clean procedure of the instrument.
作者 许达聪
出处 《实用预防医学》 CAS 2006年第3期768-769,共2页 Practical Preventive Medicine
关键词 生化分析仪 交叉污染 Β2-微球蛋白 尿素 Clinical chemistry analyzer Cross contamination β2 - microglobulim Urea
  • 相关文献

参考文献5

  • 1GU GY,GUO Q,GAO L.Effect of creatinine and total cholesterol reagent to total bile acid assay results[J].Chinese Journal of Clinical Laberatory Science,2005,23(5):361-362.
  • 2YU L.Reagent cross-contamination between the auto clinical chemistory analyzer chemistyies and it's avodance[J].Chinese Journal of Chinical Laboratory Science,2003,21 (3):168.
  • 3CHENP.Investigation of problems in clinical chemisty analyzer assay currently[J].Medical Sanitary Equipment,2001,(1):39.
  • 4YU L.Reagent cross contamination between the auto clinical chemis tory analyzer chemistyies and it's avodance[J].Chinese Journal of Chinical Laboratory Science,2003,21 (3):168.
  • 5LI H,MIAO YY,GUO ZJ.Effect of anatyze order of auto clinical chemistry malyzor to assay results[J].Chinese Jonrnal of clinical La boratory Science,2001,19(4):212.

同被引文献10

  • 1Doumas BT,Kwork-cheung PP,Perry BW,et al.Candidate Reference Method for Determination of Total Bilirubin in Sevum:Development and Validation[J].Clin Chem,1985,31:1779.
  • 2Tokuda K,Tanimoto K.The measurement of bilirubin and direct bilirubin[J].Clin Chem,1993,22(2):116.
  • 3Franzini C,Cattozzo G.Caffeine-splitting of bilirubin/albumin complex:its relevance to the spectrophotometry of bilirubin in serum[J].Clin Chem,1987,33:597-599.
  • 4Lu KC,Gooding KM,Regnier FE.Rapid analysis of bilirubin in neonatal serum.Ⅰ.The binding of bilirubin to albumin[J].Clin Chem,1979,25(9):1608-1621.
  • 5Ihara H,Nakamura H.In vitro effects of lingt on serum bilirubin subfractions measured by high-perfor mance liquid chromatography:comparison with four routine methods[J].Clin Chem,1992,38:2124-2129.
  • 6Tabb WM,Rodarte J.Urea Synthesis by Hemic Cells[J].Clin Chem,1964,10:1059-1065.
  • 7Naslund B,Stahle L.Luminometric single step urea assay using ATP-hydrolyzing urease[J].Clin Chem,1998,44:1964-1973.
  • 8于嘉屏.全自动生化分析仪及其试剂间化学污染对检测结果的影响[J].中华检验医学杂志,2007,30(11):1301-1302. 被引量:62
  • 9郑利平,李世葵,李生.β2-微球蛋白试剂对尿素测定结果的干扰及排除方法[J].国际检验医学杂志,2007,28(12):1138-1138. 被引量:1
  • 10于雷.生化自动分析仪项目间试剂的交叉污染及其避免方法[J].临床检验杂志,2003,21(3):168-168. 被引量:146