摘要
采用5-Br-PADAP-OP显色体系测定铜,考查了测定条件。以正三辛胺萃取分离干扰高子,考查了萃取分离条中。人工合成样品回收率100.8%,铝合金标样测定误差<-2.2%,人发分析结果的相对标准偏差<4.0%。
Br-PADAP-OP coloured system is used to determine copper. The meassuring conditions are researched. Trioctylamine is used to entract and seperate from interference ions. The conditions of extracting and seperating from interference ions are studied. The recovery of synthetic sample is 100. 8%. The relative error of determining results for standard samples of aluminium alloy are less than -2. 2%. The relative standard deviation of analytical results for human hair are less than 4. 0%.
出处
《光谱实验室》
CAS
CSCD
1996年第4期60-63,共4页
Chinese Journal of Spectroscopy Laboratory
关键词
显色体系
铜
萃取
分光光度法
正三辛胺
Br-PADAP-OP Coloured System, Copper, Aluminium Alloy, Human Hair,Extraction-Spectrophotometry.