摘要 小型氡室是一种综合测氡仪器的标定装置,适合目前各种测氡仪器的标定。湿度、温度、闪烁探测器的漏计现象等因素影响氡室内氡浓度。湿度影响着气溶胶的密度,湿度越大,气溶胶的密度也越大,导致探测器探测的计数也变大;温度增加,分子的扩散系数随着增加,使镭源释放的氡量增加,导致氡浓度变大;闪烁体的发光衰减时间较长,测量大活度样品时,会出现漏计现象。对应不同的影响因素采取了相应的修正措施。 Small-scale radon chamber is one comprehensive calibrating instrument, which is fit to various kinds of radon detectors at present. Humidity, temperature, missing counting phenomenon of scintillation detector, etc, influence accuracy of radon concentration of radon chamber. The humidity is influencing the density of the aerosol. The bigger the humidity is, the larger the density of the aerosol is, so detector counts more; The diffusion coefficient of the molecule increases with temperature increasing, which makes radon released from radium source increase and causes radon Concentration to increase; Decay time of scintillation luminescence is relatively long. While measure big activity sample, missing counting phenomenon exists. This paper discusses these factors. According to different influencing factors, corresponding revision measurements should be taken.
机构地区 东华理工学院
出处 《东华理工学院学报》 2006年第2期137-139,共3页 Journal of East China Institute of Technology
关键词 准确性 氡浓度 湿度 温度 漏计 accuracy Radon concentration humidity temperature missing counting
  • 相关文献

参考文献3

二级参考文献3

共引文献4

同被引文献8

  • 1张震寰,赵晶,刘岩.氡浓度标准装置的质量控制及保证[J].计测技术,2007,27(2):33-36. 被引量:2
  • 2J. BIGU, The design of a radon/thoron calibration facility [R], ELLIOT LAKE LABORATORY, Canada Centre for Mineral and Energy Technology, Mineral and Energy Technology Mining Research Laboratories Division Re-port MRP/MRL 81-46, 1981.
  • 3BIGU J, A preview of the national radon/thoron test facility[R], ELLIOT LAKE LABORATORY, Canada Centre for Mineral and Energy Technology, Mineral and Energy Technology Mining Research Laboratories Division Re-port M&ET/MRL 87, 1986.
  • 4金以惠.过程控制[M].北京:清华大学出版社,2000.
  • 5BIGU J. The design of a radort/thoron calibration facility, Elliot lake laboratory, Canada Centre for Mineral and Energy Technology [ M ]. Mineral and Energy Technology Mining Research Laboratories Division Re - port MRP/MRL 81 -46, 1981.
  • 6BIGU J, A preview of the national radon/thoron test facility, ELLIOT LAKE LABORATORY, Canada Centre for Mineral and Energy Technology [ M ]. Mineral and Energy Technology Mining Research Laboratories Division Re-port M&ET/MRL 87, 1986.
  • 7张军,周书民.基于PCI-8503的氡室监控系统研究[J].微计算机信息,2007,23(19):118-119. 被引量:2
  • 8周伟,方方,郑永明,阎萍,刘易.氡标准室中浓度智能控制系统的实现[J].核电子学与探测技术,2007,27(4):713-715. 被引量:3

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部