摘要
将日立110XF型数字胃肠机中的SX-YB2型悬吊摄影系统从主机中分离出来,再与国产FSK302-1及FSB302-1主控高压系统和F78-III型平床摄影系统相配合,组成新的大型综合X线摄影系统。主要将灯丝加热电路、旋转阳极启动运转电路、X线发生控制电路进行较大完善改造,并增加了部分电源电路及灯丝加热电路。
The new general photography system is composed of overhang of SX-YB2 departed from the mainframe of HITACHI Digital Fluoroscpic (model 110XF), topmanagement system of FSK302-1 and FSB302-1 and leveled adiogaphy system of F78-Ⅲ. The filament heater circuit, rotating anode startup and running circui and X-ray produce circuit are mainly changed, and some circuits in power and filament heat are added.
出处
《医疗卫生装备》
CAS
2006年第6期77-77,79,共2页
Chinese Medical Equipment Journal
关键词
日立
国产
X线机
改装
HITACHI
indigenous X-ray
repark