管电流调制技术在多层螺旋CT胸部检查中的应用评价 被引量:7

Attenuation-based online tube current modulation applied to multislice CT of the chest
下载PDF
导出
摘要 目的评价管电流调制技术降低多层螺旋CT(MSCT)胸部扫描剂量的程度,以及对CT图像质量的影响。方法将100例受检者随机等分成两组(管电流调制组和管电流恒定组)。分别统计CT剂量权重指数(CTDIw)、剂量长度乘积(DLP)及有效毫安秒(mAs)。图像质量分为优、良、差三级进行评判,并记录评价结果。数据资料采用成组t检验和秩和检验进行统计分析。结果管电流调制组有效mAs为(87.69±2.74)mAs,较恒定组(100mAs)降低约12%(P<0.001)。管电流调制组CTDIw(6.16±0.23mGy),明显低于恒定组(7.0mGy)(P<0.001)。管电流调制组DLP(202.03±18.27mGy·cm)约为恒定组(230.13mGy·cm)的87.7%(P<0.001)。相同扫描条件和图像后处理方式下,管电流调制组的图像质量不仅未降低,而且还可选择性地改善(P>0.05)。结论管电流调制技术可有效降低CT剂量,并不牺牲图像质量,有较好的临床应用前景。 Objective To evaluate the radiation dose saved by attenuation-based online tube current modulation applied to multislice CT (MSCT) of the chest, and impact on the quality of CT images. Methods One hundred patients who underwent MSCT of the chest were classified randomly two groups (attenuation-based online tube current modulation group and constant tube current). Consecutive acquisitions including weighted-CT-dose index (CTDWI), dose long product (DLP) and effective mAs were obtained in each patient and compared for each group. Three experienced chest radiologists who were unaware of the CT technique reviewed CT images for overall image quality using a 3-grade scale (excellent, good and worst). The data were analyzed using the unpaired t test and Wilcoxon's signed rank test. Results Mean effective mAs (87.69±2.74 mAs) reduction was 12% for the chest scan in attenuation-based online tube current modulation group. CTDWI (6.16 ± 0.23 mGy) and DLP (202.03±18.27 mGy ·cm) reduction were obviously less than the constant tube current group. The differences were statistically significant (P〈0. 001). The assessment of image quality showed no significant differences under same protocol and post-processing technique (Wilcoxon's signed rank test, P〉0.05). Conclusion Attenuation-based online tube current modulation used with MSCT should be considered as a promising technique because it saves as much as 13% of the radiation dose on chest scan and don't impair the image quality.
出处 《中国医学影像技术》 CSCD 北大核心 2006年第6期858-860,共3页 Chinese Journal of Medical Imaging Technology
基金 教育部留学回国人员科研基金资助(2002247)。
关键词 管电流调制 体层摄影术 X线计算机 胸部 Attenuation-based online tube current modulation Tomography, X-ray computed Chest
  • 相关文献

参考文献8

  • 1Moran LM,Rodriguze R,Calzado A,et al.Image quality and dose evaluation in spiral chest CT examinations of patients with lung carcinoma[J].Br J Radiol,2004,77(922):839-846.
  • 2Mayo JR,Aldrich J,Muller NL.Radiation exposure at chest CT:a statement of the fleischner society[J].Radiology,2003,228(1):15-21.
  • 3Tack D,Widelec J,De Maertelaer V,et al.Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic cinusitis[J].AJR,2003,181(4):939-944.
  • 4Hagtvedt T,Aalokken T,Notthellen J,et al.A new low-dose CT examination compared with standard-dose CT in the diagnosis of acute sinusitis[J].Eur Radiol,2003,13(5):976-980.
  • 5Dinkel HP,Sonnenschein M,Hoppe H,et al.Low-dose multislice CT of the thorax in follow-up of malignant lymphoma and extrapulmonary primary tumors[J].Enr Radiol,2003,13 (6):1241-1249.
  • 6Tack D,De Maertelaer V,Gevenois PA.Dose reduction in multidetector CT using attenuation-based online tube current modulation[J].AJR,2003,181(2):331-334.
  • 7Kalender WA,Wolf H,Suess C,et al.Dose reduction in CT by on-line tube current control:principles and validation on phantoms and cadavers[J].Eur Radiol,1999,9(2):323-328.
  • 8Gies M,Kalender WA,Wolf H,et al.Dose reduction in CT by anatomically adapted tube current modulation.I.Simulation studies[J].Med Phys,1999,26(11):2235-2247.

同被引文献104

引证文献7

二级引证文献46

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部