多层螺旋CT轴位扫描图像与MPR图像空间分辨率和低对比度分辨率的比较 被引量:2

Comparison of the spatial resolution and low contrast resolution between the axes location image and MPR image of multi-slice helical CT
下载PDF
导出
摘要 目的本文通过实验对多层螺旋CT轴位扫描图像与Z轴MPR图像的空间分辨率和低对比度分辨率进行了定量分析比较。方法通过分别将Catphan500模体中CTP538与CTP515模块以不同方向浸入自制同等密度的葡萄糖溶液,在MDCT下进行扫描。结果在相同扫描的参数下,轴位扫描图像的空间分辨率和低对比度分辨率比对应的MPR图像好,但相差并不显著,且选择薄层扫描可以使MPR图像更接近轴位扫描图像。结论结合临床的特点,本文提出合理地使用MPR技术的建议。 Objectives In this paper a test was carried on to analyze quantitatively the spatial resolution and low contrast resolution of both the normal image and MPR image. Methods From perpendicularly directions immersed the CTP538 and CTPS15 modules of CatphanS00 in confected dextrose liquor which density was identity with the modules CTPS15 we scanned them in MDCT respectively. Result It was indicated that under the same scanning conditions both of the resolutions and low contrast resolution of the axes location image is superior to that of MPR image. But the difference is not very notable. The thinner scanning slice be used, the smaller difference between the two types of images. Conclusions In the end of this paper, some suggestions on the clinical using of MPR were put forward.
出处 《医疗设备信息》 2006年第6期84-87,共4页 Information of Medical Equipment
关键词 MPR图像 扫描层厚 重建层厚 空间分辨率 低对比度分辨率 MPR image scanning slice reconstructed slice spatial resolution low contrast resolution
  • 相关文献

参考文献6

二级参考文献18

  • 1柳澄,秦维昌.多层螺旋CT(一)[J].医学影像学杂志,2000,10(3):194-195. 被引量:42
  • 2Stanford W (ed). Cardiac imaging, part Ⅰ and Ⅱ[J].J Thorac Imaging ,2000,15(4) and 2001; 16(1).
  • 3Rubin GD. Techniques for performing multidetector-row computed tomographic angiography[J]. Tech Vasc Interv Radiol 2001, (4):2- 14.
  • 4Wang G, Vannier MW .Spatial variation of section sensitivity profile in spiral computed tomography[J].Med Phys, 1994,21 (9) : 1491 - 75.
  • 5McCollough CH, Kanal KM, Lannutti N, et al.Experimental determination of section sensitivity profiles and image noise in electron beam computed tomography[J]. Med Phys,1999,26(2):287-95.
  • 6North American symptomatic carotid endartetectomy trail collaborators.benificial effect of carotid endartetectomy in symptomatic patients with high grade carotid stenosis[J] .N Engl J Med,1991,325:445-453.
  • 7BrinkJA.Technical aspects of helical(spiral)CT[J].Radiol Clin North Am,1995,33:341.
  • 8Dillon EH .Vanleeuwen MS,Femandez MA,et al.Spiral CT angiography[J].AJR,1993,160:1273.
  • 9Neri E ,Caramella D,Falaschi F,et al.Virtual CT intravascular endoscopy of the aorta:pierced surface and floating shape thresholding artifacts[J].Radiolody,1999,212:276 - 279.
  • 10Manninen HJ,Rasanen H,vanninen RL,et al.Human carotid arteries:correlation of intravascular US with angiographic and histopathologic findings[J].Radioiogy,1998,206:65 - 74.

共引文献12

同被引文献32

  • 1兰永胜.200例皮层下动脉硬化性脑病的CT诊断分析[J].西南军医,2005,7(6):37-38. 被引量:1
  • 2余晓锷,占杰,李萍,李婵娟.统计学方法在CT低对比度分辨力自动检测中的应用[J].第四军医大学学报,2006,27(11):978-979. 被引量:3
  • 3柳澄.充分发挥各向同性扫描的优势,开拓多层CT新的应用领域[J].医学影像学杂志,2007,17(1):1-3. 被引量:32
  • 4Verdun FRD, Ott JG, Tapiovaara M J, et al. Image quality in CT:From physical measurements to model observers[J]. Phys Med, 2015,31 (8). 823- 843.
  • 5Hernandez-Gironl Geleijns J,Calzado A,el at. Automated assessment of low contrast sensitivity for CT systems using a model observer[J].Med Phys. 2011,38(Suppl 1) .$25-$35.
  • 6Verdun FR,Denys A,Valley JF,el at. Detection of low-contrast objects:experimental comparison of single-and multi-detector row CT with a phantom[J]. Radiology, 2002,223(2) .426-431.
  • 7Hernandez-Giron I, Calzado A, Geleijns J,et al. Low contrast detectability performance of model observers based on CT phantom images:kVp influence[J]. Phys Med, 2015,31(7): 798-807.
  • 8Solomon J,Samei E.Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms:FBP and SAFIRE[J}. Med Phys, 2014,41 (9): 091908.
  • 9Hernandez-Giron I,Calzado A,Geleijns J,et al. Comparison between human and model observer performance in low-contrast detection tasks in CT images:application to images reconstructed with filtered back projection and iterative algorithms[J].Br J Radiol, 2014,87(1039) :20140014.
  • 10yon Falck C,Rodt T,Waldeck S,et al.A systematic approach towards the objective evaluation of low-contrast performance in MDCT:combination of a full reference image fidelity metric and a software phantom[J].Eur J Radiol,2012,81(ll):3166-3171.

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部