摘要 采用CFD仿真技术,结合所设计的微型轴流血泵,对血泵叶轮中的速度、压力、应力等分布进行了仿真分析。研究结果表明:血泵中的流体具有非常复杂的流动情况,为了避免流动分离、压力变化过大等情况,减少剪切应力,对血泵叶轮结构提出了相应的改进意见。 CFD technology has been used in the design of minimal axial blood pump. The divisions of velocity, pressure, and stress on the impeller of blood pump has been simulated and analyzed. The research results indicate: The blood flowing state in the blood pump is very complex. Several betterments on the structure of the pump impeller have been proposed, which can be used to avoid separating flow, to avoid, large changing of pressure, to reduce shear stress.
出处 《机械》 2006年第7期4-6,共3页 Machinery
基金 国家自然科学基金(50275151) 湖南省自然科学基金(02JJY2080)
关键词 血泵 仿真 叶轮结构 blood pump: simulation: impeller structure
  • 相关文献

参考文献6

  • 1Weiss WJ. Recent improvement in a completely implanted total artificial heart[J]. ASAIO Journal, 1996, 42: 342.
  • 2Nose Y. Can we develop a total implantable rotary blood pump[J].Artificial Organs, 1995, 19 ( 7 ) : 561.
  • 3Kawahito K, Damm G, Benkowski R. et al. Ex vivo phase 1 evalution of the DeBakey/NASA axial flow ventricular assist device[J]. Artificial Organ, 1996, 20 (1) : 47.
  • 4NiimiH.切变流动中红细胞膜上的交变载荷:溶血的潜在原因[J].生物医学工程国外分册,1986,9(4):292-296.
  • 5陈恩,俞晓青,丁文祥.轴流泵的设计与实验研究进展[J].生物医学工程学杂志,2000,17(4):469-472. 被引量:11
  • 6苏明德,黄素逸.计算流体力学基础[M].北京:清华大学出版社,1997.

二级参考文献10

  • 1Sukumar R,Athavake MM,Makhijani VB et al.Application of computational fluid dynamics techniques to blood pumps[].Artificial Organs.1996
  • 2Damm G,Mizuguchi K,Aber G et al.Axial flow ventricular assist device: System performance considerations[].Artificial Organs.1994
  • 3Kawahito K,Damm G,Benkowski R et al.Ex vivo phase 1 evaluation of the DeBakey/NASA axial flow ventricular assist device[].Artificial Organs.1996
  • 4Reddy VM,Liddicoat JR,klein JR et al.Fetal cardiac bypass using an in line axial flow pump to minimize extracorporeal surface and avoid priming volume[].The Annals of Thoracic Surgery.1996
  • 5Sess T,Meyns B,Spielvogel et al.Hemodynamic system analysis of intraarterial microaxial pumps in vitro and in vivo[].Artificial Organs.1996
  • 6Anai H,Wakisaka Y,Nakatani T et al.Relationship between pump speed design and hemolysis in an axial flow blood pump[].Artificial Organs.1996
  • 7Shimono T,Makinochi K,Nose′Y.Total erythrocyte destruction time:the new index for the hemolytic performance of rotary blood pumps[].Artificial Organs.1995
  • 8Ymazaki K,Umezu M,Koyanaki H et al.A miniature intraventricular axial flow blood pump that is introduced through left ventricular apex[].ASAIO Journal.1992
  • 9Antaki JF,Butler KC,Kormos RL et al.In vivo evaluation of the Nimbus axial flow ventricular assist system criteria and methods[].ASAIO Journal.1993
  • 10Nose’Y,Kawahito K,Nakazawa T.Can we develop a nonpulsatile permanent rotary blood pump?Yes, we can[].Artificial Organs.1996