摘要 目的:评价各种镀膜钕铁硼磁体的细胞相容性。方法:将镀镍、氮化钛、银、铜的钕铁硼磁体、未镀膜磁体及托槽分别制备浸提液,与原代培养大鼠牙龈成纤维细胞共同培养,通过A lam arB lueTM法测细胞活性及扫描电镜观察细胞形态的改变。结果:5%托槽、镀氮化钛磁体组的A lam arB lue的还原率分别为74.56%±1.60%、76.54%±3.82%,对照组为79.16%±3.17%,无显著性差异(P>0.05);10%托槽、镀氮化钛磁体组的A lam arB lue的还原率分别为76.00%±2.17%、72.86%±2.47%,对照组为77.76%±2.76%,无显著性差异(P>0.05)。其余各组与对照组相比,差异有显著性(P<0.05)。加入托槽、镀氮化钛磁体浸出液的培养细胞表面形态良好、细胞膜完整,镀镍磁体浸出液的培养细胞表面形态尚好,镀铜及未镀膜磁体浸出液的培养细胞表面形态偶见不光滑、不完整。结论:镀氮化钛的钕铁硼磁体具有良好的细胞相容性。 Objective: To investigate the cell biocompatibility of different coating NdFeB magnets. Methods: Nickel, titanium-nitride, silver, or copper Was coated on NdFeB magnets. The coated magnets, bare magnets and brackets were incubated with DMEM. AlamarBlue TM and scan electron microscope were used to examine the changes of cell activity of primary rat gingival fibroblast cell cultured with those prepared DMEM. Results: The percentage of reduced AlamarBlue in 5% brackets, titanium nitride magnets groups was 74. 56%± 1. 60%, 76. 54%± 3. 82 %, there was no significant difference compared with the control group (79.16 % ± 3. 17% ) (P〉0.05);The percentage of reduced AlamarBlue in 10% brackets, titanium nitride magnets groups was 76.00%±2.17% ,72.86%±2.47 % ,there was no significantly difference with control group (77.760% ± 2. 76 %) (P 〉0. 05) ; The percentage of reduced AlamarBlue in other groups was significantly lower than control group (P〉0. 05) . In the bracket and titanium magnet groups the cells had smooth surface and complete membrane. The cell surface in nickel group was fine. The cells in copper and bare magnet groups had coarse surface and incontinuous membrane. Conclusions: Magnet coated titanium nitride has good biocompatibility.
出处 《陕西医学杂志》 CAS 北大核心 2006年第8期919-921,共3页 Shaanxi Medical Journal
基金 辽宁省教育厅攻关计划资助项目(项目编号:202013180)
关键词 细胞 生物相容性材料 @钕铁硼磁体 Cell Biocompatile materials @NdfeB magnet
  • 相关文献

参考文献5

二级参考文献14

  • 1Sekiguchi y. Applications of shape memory alloys: Medical applications. In : Funakubo H, editor. Shape memory alloys [ M ]. New York : Gordon and Breach Science Publishers, 1987 ; 226 - 269.
  • 2Uo M, Watari F, Yokoyama A, Matsuno H, Kawasaki T. Dissolution of nickel and tissure response observed by X-ray scanning analytical microscopy [J]. Biomaterials, 1999,20(7) :747 - 755.
  • 3Grill A. Diamond-like carbon coatings as biocompatible materials-an overview [ J ]. Diamond and related materials, 2003, 12 ( 1 ) :166- 170.
  • 4Peltonen L. Nickel sensitity in general population contact dermatitis[J]. Biomaterials 1979,5( 1 ) :27 - 32.
  • 5Thomsen P, Ericson LE. Inflammatory cell response to bone implant surfacs, in The Bone Biomaterial Interface [ C ]. J E Davies ( ed ),University of Toronto Press,Toronto, 1991 ; 135 - 164.
  • 6Fergouson A B. The ionization of metal implants in living tissues[J]. J Bone Joint Surg(Am), 1960,42( 1 ) :77 - 9o.
  • 7Jones MI, McColl R, Grant DM, Parker KG, Parker TL. Protein adsorption and platelet attachment and activation, on TiN, TiC, and DLC coatings on titanium for cardiovascular applications [ J ]. J Biomed Mater Res,2000,52(4) :413 - 421.
  • 8Kapanen A, Ryhanen J, Danilov A, Tuukkanen J. Effect of nickeltitanium shape momory metal alloy on bone formation [ J ].Biomaterials, 2001,22 ( 11 ) : 2475 - 2480.
  • 9Wever D J, Veldhuizen AG, de Vries J, Busscher H J, Uggs DRA, van Horn JR. Electrochemical and surface characterization of a nickeltitanium alloy[J]. Biomaterials, 1998,19(7) :761 - 769.
  • 10Wataha JC, Lockwood PE, Marek M, Ghazi M, Ability of Nicontaining biomedical alloys to activate monocytes and endothelial cells in vitro[J] .J biomed Mater Res,1999(2) ,45:251 - 257.