鼻咽癌调强放疗中的摆位误差 被引量:16

Set-up Error in Intensity Modulated Radiotherapy for Nasopharyngeal Carcinoma
下载PDF
导出
摘要 [目的]用电子射野图像器件(EPID)拍摄的射野片研究调强放疗治疗鼻咽癌过程中的摆位误差。[方法]比较8例接受调强放疗的初治鼻咽癌患者的数字重建影像(DRR)和EPID图像,得到各个骨性标志间各个方向的偏差,分别以左右、前后和头脚方向上的最大偏差代表该方向的摆位误差。[结果]3个方向的摆位误差范围是-5mm到5.5mm,平均值是(-0.87±1.3)mm,(-0.28±1.5)mm和(-0.55±1.6)mm,大于2mm的误差分别占17.3%、14.3%和17.3%,且各次治疗的摆位误差之间无显著性差异。[结论]鼻咽癌调强放疗的误差在可以接受的范围,而且各次治疗间也没有显著差异,首次治疗时拍摄射野片验证是非常重要的。 [Purpose]To investigate the set-up error in intensity modulated radiotherapy (LMRT)with electric portal image device (EPID)in nasopharyngeal carcinoma(NPC).[Methods]Diserepancies of bone an atomy markers in various directions were achieved by comparison of digifally reconstructed radiograph(DRR)with image of EPID in 8 cases of NPC initial treated with IMRT.The maxium discrepancies in mediolateral,antero-posterior and cranio-caudal directions were defined as the set-up error of each direction.[Results]The range of setup error in three directions was -5mm to 5.5mm.The mean set-up error in three directions was (-0.87±1.3)mm,(-0.28±1.5)mm and (-0.55±1.6)mm respectively,and errors larger than 2mm accounted for 17.3%,14.3%and 17.3% respectively.No significant difference was found in set-up error among each fraction ations.[Conclusion]The set-up error in IMRT for NPC is acceptable,and also no significant difference in every fractionations.Portal verified image in the first treatment is very important.
机构地区 江苏省肿瘤医院
出处 《肿瘤学杂志》 CAS 2006年第4期269-271,共3页 Journal of Chinese Oncology
基金 江苏省135工程重点学科-肿瘤放射治疗学和"鼻咽癌调强放射治疗的研究"课题资助
关键词 鼻咽肿瘤 调强放疗 摆位误差 nasopharyngeal neoplasms intensity modulated radiotherapy det-up error
  • 相关文献

参考文献8

  • 1Kam MK, Chau RM, Suen J, et al. Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalation [J]. Int J Radiat Oncol Biol Phys, 2003, 56(1):145-157.
  • 2翟振宇,何侠,魏青,张宜勤.鼻咽癌调强放射治疗的剂量学特点[J].肿瘤学杂志,2005,11(5):337-340. 被引量:21
  • 3van Lin EN, van der Vight L, Huizenga H, et al. Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customised head and neck support[J]. Radiother Oncol, 2003,68(2):137-148.
  • 4林承光,邓小武,黄峻,温志祥,肖达宜,林览峰.鼻咽癌适形放疗实施过程精确度和重复性研究[J].肿瘤学杂志,2004,10(4):208-210. 被引量:32
  • 5Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability[J]. Int J Radiat Oncol Biol Phys, 1999,43:905-919.
  • 6何侠,翟振宇,纪红,张宜勤.初治鼻咽癌调强适形放疗近期临床观察[J].肿瘤防治研究,2006,33(4):272-274. 被引量:31
  • 7Schallenkamp JM, Herman MG, Kruse JJ, et al. Prostate position relative to pelvic bony anatomy based on intraprostatic gold markers and electronic portal imaging [J].Int J Radiat Oncol Biol Phys, 2003, 63:800-811.
  • 8郑步宏,潘建基,徐鹭英,陈传本,张瑜,张秀春.垫内置标记点法测量鼻咽癌适形放疗摆位误差的研究[J].中华放射肿瘤学杂志,2005,14(3):218-221. 被引量:19

二级参考文献25

  • 1[2]Kinxie JJ, Hanks GE, Maclean, CJ et al. Patterns of care study: Hodgkin's Disease relapse rates and adequacy of portals[J]. Cancer, 1983, 52:2223-2226.
  • 2ICRU. Rescribing, recording, and reporting photon beam therapy. ICRU Report 50. Bethesda: ICRU, 1993.
  • 3Gilbeau L, Octave-Prignot M, Loncol T, et al. Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors. Radiother Oncol,2001, 58:155-162.
  • 4Zhu Y, Stovall J, Butler L, et al. Comparison of two immobilization techniques using portal film and digitally reconstructed radiographs for pediatric patients with brain tumors. Int J Radiat Oncol Biol Phys,2000,48:1233-1240.
  • 5Marc W. Munter, et al. Evaluation Of Salivary Gland Function After Treatment Of Head-And-Neck Tumors With Intensity-Modulated Radiotherapy By Quantitative Pertechnetate Scintigraphy[J]. Int J Radiation Oncology Biol Phys, 2004, 58(1): 175-184.
  • 6Ching-Yeh Hsiung, et al. Attenuation of Radiation Dose by the Skull Base Bone in Patients with Nasopharyngeal Carcinoma: Clinical Importance[J]. Radiology, 200l, 218: 457-463.
  • 7Michael K. M. Kam, et al. Intensity-Modulated Radiotherapy In Nasopharyngeal Carcinoma: Dosimetric Advantage Over Conventional Plans And Feasibility Of Dose Escalation[J]. Int J Radiation Ontology Biol Phys, 2003, 56( 1 ): 145-157.
  • 8Marc W, Munter, et al. Evaluation Of Salivary Gland Function After Treatment Of Head-And-Neck Tumors With Intensity-Modulated Radiotherapy By Quantitative Pertechnetate Scintigraphy[J]. Int J Radiation Ontology Biol Phys, 2004,58(1): 175-184.
  • 9Bin Steh, Shiao Ywoo, E Brian Bueler. Intensitymodulated radiation therapy(IMRT):a new promising technology in radiationeology[J]. The Oneologist, 1999,4:433-442.
  • 10Emarni B, Lyrnan J, Brown A, et al. Tolerance of normal tissue to the rapeutieir radiation[J]. Int J Radit Oneol Biol Phys,1991,21:109-122.

共引文献94

同被引文献75

引证文献16

二级引证文献91

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部