期刊文献+

C8051F350在热药灌注机中的应用 被引量:7

Application of C8051F350 in hot medicine filler
下载PDF
导出
摘要 热药以及病变组织温度的精确测量对于热药灌注机的疗效至关重要.以Pt100作为测温传感器,采用SoC C8051F350作为热药灌注机测温系统的核心控制芯片,实现热疗过程中温度的精确测量控制,在保证了测量精度要求的同时又减小了产品体积,在0~50℃范围内,其绝对温度精度可达±0.2℃. It is very important that the precise temperature of hot medicine liquid and pathological organizations must be measured in hot medicine filling process. By using Pt100 as a temperature sensor and SoC C8051 F350 as a core chip, the precious measure and control for the temperature in hot medicine liquid treatment are realized when the requested measure accuracy is assure and the volume of the products is reducesd. The absolute accuracy can reaches ±0.2 ℃ at the range of from 0 ℃ to50 ℃.
作者 谷兆麟 赵杰
出处 《天津工业大学学报》 CAS 2006年第4期14-17,共4页 Journal of Tiangong University
基金 中国教育部科学技术重点项目(204010)
关键词 热药灌注 温度测量 SOC PT100 hot medicine filling temperature measurement SoC Pt100
  • 相关文献

参考文献5

  • 1潘琢金 孙德龙 夏秀峰译.C8051F单片机应用解析[M].北京:北京航空航天大学出版社,2002..
  • 2吴水才,白燕萍,南群,夏雅琴.癌热疗中超声无创测温方法的研究[J].国外医学(生物医学工程分册),2002,25(1):35-38. 被引量:14
  • 3BURD R,DZIEDZIC TS,XU Y,et al.Tumor cell apoptosis,lymphocyte recruitment and tumor vascular changes are induced by low temperature,long duration (fever-like)whole body hyperthermia[J].J Cell Physiol,1998,177(1):137 -147.
  • 4WANG X Y,OSTBERG J R,REPASKY E A.Effect of fever-like whole-body hyperthermia on lymphocyte spectrin distribution,protein kinase C activity,and uropod formation[J].J Immunol,1999,162(6):3 378 -3 387.
  • 5王俊松.基于BP神经网络的温度传感器动态建模及自适应标定[C]//中国自动化学会控制理论专业委员会.第24届中国控制会议论文集.广州:华南理工大学,2005:1 624-1 626.

二级参考文献18

  • 1Meaney P,Paulsen K,Hartov A,et al. Microwave imaging for tissue assessment :Initial evaluation in multitarget tissue-equivalent phantoms [J]. IEEE Trans BME, 1996,43(9): 878-890.
  • 2Diederich CJ,Nau WH,Deardorff DL,et al. MR thermal monitroring of ultrasound interstitial thermal therapy[J]. Proceedings of SPIE-The International Society for Optical Engineering, 1999,3594:178-184.
  • 3Paulsen K,Moskowitz M,Ryan T,et al. Initial in vivo experience with EIT as a thermal estimator during hyperthermia [J]. Int J Hyperthermia, 1996, 12 (5): 573-591.
  • 4Lu BY,Lin WL,Yang RS,et al. Temperature distributions of bone tumor therapy using scanned focused ultrasound system. Ann Int Conf IEEE Eng Med Biol,Atlanta, GA, USA. 1999,2:1272.
  • 5Singh VR,Yadav S. Non-invasive measurement of temperature in ultrasonic hyperthermia[C]. IEEE Engineering in Medicine & Biology Society 10th annual International Conference, New Delhi, India, 1998. 0850-0851.
  • 6Ueno S.Ultrasound thermometry in hyperthermia[C].Proc IEEE Ultrasonics symposium,Kawasaki,Japan.1990.1645-1692.
  • 7Seip R,Ebbini ES,Donnell MO,et al.Non-invasive detection of thermal effects due to highly focused ultrasonic fields[C].Proc IEEE Ultrasonics symposium,Piscataway,NJ,USA.1993.1229-1232.
  • 8Gertner MR,Worthington AE,Wilson BC,et al.Ultrasound imaging of thermal therapy in vitro liver[J].Ultrasound in Medicine and Biology,1998,24(7):1023-1032.
  • 9Fukukita H,Ueno SI,Furuya N,et al. Noninvasive temperature measurement using ultrasound nonlinear effect[C].Ultrasound nonlinear effect,Kawasaki,Japan.1990.663-665.
  • 10Sato T,Ichida N.Real time non-invasive abservation of change of temperature destribution using ultrasonic non-linear parameter[J].Automedica,1987,8:283-294

共引文献71

同被引文献10

引证文献7

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部