摘要
目的研究全自动分析CK和CK-MB试剂对镁测定的影响。方法通过混合血清或蒸馏水标本单独测镁与先测CK或CK-MB,再测镁的结果比较来讨论。结果单独测镁与先测CK或CK-MB,再测镁的结果存在差异显著性性。结论CK和CK-MB试剂因含高浓度镁,对镁测定有正干扰。
Objective To study the effect of CK and CK-MB reagent on detecting magnesium by an automatical analyzer. Methods Determination of magnesium was carried out alone,compared with it was done after measuring CK or CK MB. Results Significant differences were observed between the mean values of magnesium concentration in two groups Conclusion CK and CK-MB had significant influence on detecting magnesium, because they contented the high concentration of magnesium.
出处
《实用医技杂志》
2006年第16期2805-2806,共2页
Journal of Practical Medical Techniques
关键词
干扰
全自动分析
镁
Interference
Automatical analysis
Magnesium