降低毫安秒对新生儿头颅CT质量的影响 被引量:8

The Effect of Decreasing mAs on Image Qulity of the Newborn Brain.
下载PDF
导出
摘要 目的探讨新生儿头颅CT扫描中降低mAs与图像质量的关系及低剂量CT扫描的可行性。方法将80例新生儿头颅CT检查随机分为三组进行前瞻性研究,其中100mAs组35例,225mAs组21例,260mAs组24例。由两名有经验的放射医师以双盲的方法使用4分制评分,对三组图像中头部六个部位的解剖结构的清晰度进行评分,说明CT图像质量的优良程度。使用方差分析检查三组图像质量评分有无显著性差异。结果所有图像质量均能满足诊断要求。2个阅读者对三组不同毫安秒图像中1~6解剖部位的总评分差异均无显著性,对每一个解剖部位的评分中,260mAs组及225mAs组稍高于100mAs组,但三组中两两比较无显著性差异(P〉0.05)。结论应用100mAs低剂量新生儿头颅CT扫描是可行的,图像质量能够满足临床要求。 Objective To determine wheather a lower radiation dose technique could be used in CT of the newborn brain without jeopardising the diagnostic accuracy of the images. Methods A group of 24 newborn underwent CT using 260 mAs, 21under scanning at 225mAs, 35 underwent scanning at 100 mAs, Anatomy details and the confidence level in reaching a diagnosis were evaluated by two radiologists in a double - blinded manner using a 4 - point scoring system. Results For both reader there was no statistically significant difference between three group total score of 1~6 anatomical detail and each of 6 anatomical detail although score for each of 6 anatomical detail in 260 mAs group and 225mAs group was a little higher than 100mAs group. Condution Radiation dose reduction in newborn brain CT is feassible in clinical use.
出处 《中国辐射卫生》 北大核心 2006年第3期277-278,共2页 Chinese Journal of Radiological Health
关键词 低剂量CT 新生儿 头颅 体层摄影术 X射线计算机 Low - dose CT Newborn Brain Tomography, X - ray Coputed
  • 相关文献

参考文献11

  • 1Cohnen M, Fisher H, Mamacher J, et al. CT of the head by use of reduced current Kilovoltage: relalionship between image qualily and dose reduction[J]. AJNR, 2000,21: 1654- 1660.
  • 2Chan CY, Wong YC, Chau LF, et al. Radiation dose reduction in pediatrics cranial CT[ J ]. Pediatric Radial, 1999,29 : 770 - 775.
  • 3Calzdo A, Rodriguez R, Murloz A, et al. Quality criteria implementation for brain and lumber spine CT examinations[J].Bri J Radiol, 2000,73384 - 73395.
  • 4Haaga JR. Radiation dose Management: Weighling Risk Versus Benefit[J]. AIR 2001,177:289 - 291.
  • 5Paterson A, Fresh DP, Donnelly LF. Helical CF of the body are settings . adjuster for pediatric patients [J]. AJR, 2001,176 : 297- 301.
  • 6Lucaya J, Piqueras J, Pena PG, et al. low - dose high - Resolution CT of the chest in Children and yang adults: dose, cooperation, artifact incidence, and image quality[ J] . ASR , 2000, 175 :985 - 992.
  • 7Rogalla P, storer B scheer Z, et al. Low - dose spiral CF: applicability to pediatric chest imaging[J]. Pediatric Radial, 1998,28 : 565 - 569.
  • 8Frush DP, Sluok CC, Holingsworth CL, et al. computer- simulated radiation dose reduction for abdominal multidetector CT of Pediatric Patients[J].AJR., 2002, 179:1107 - 1113.
  • 9Feuron T, Vucich J. Normalized pediatric organ - absorbed dose from CT examination[J] . AJR, 1987,148:171 - 174.
  • 10Cfisty M, Active bone marrow distribution as a function of age in humans[J]. Phys Med Biol, 1981,26:389-400.

同被引文献71

引证文献8

二级引证文献41

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部