心电图调制电流对于容积CT心脏扫描图像质量及曝光剂量影响的评估 被引量:45

Evaluation on the influence of electrocardiograph modulated milliampere on image quality and exposure dosage of volume CT heart scan
原文传递
导出
摘要 目的评估容积 CT(VCT)心脏扫描时,开启或关闭心电图(ECG)调制电流(mA)技术对图像质量是否有影响,并判断 ECG 调制 mA 技术是否能有效降低曝光剂量。方法将跳动心脏模具设定为高、中、低不同速率的稳定心率,采用与不同速率稳定心率相匹配的重组算法,开启 ECG 调制 mA,对心脏模具进行扫描,扫描同时记录各扫描序列的曝光剂量,在采集到的图像中选取结构均匀的部位,测量该部位的 CT 值及不同部位 CT 值的差值,并记录图像测量部位的标准差(s)值;关闭ECG 调制 mA,重复上述扫描并记录曝光剂量,对采集到的图像进行相同测量和记录。将2次扫描所测得的数据进行统计学分析。结果在同样心率条件下,ECG 调制开启时比关闭时曝光剂量有明显减低,在每分钟搏动数(bpm)为50、60、75和95的4种心率条件下,曝光剂量分别减少了48.0%、44.0%、37.0%和37.7%;2组图像的数据分析差异无统计学意义(P>0.05)。结论 VCT 心脏扫描时,开启 ECG 调制 mA 技术能够有效降低曝光剂量,同时对图像质量无明显影响。 Objective To find out whether the use of ECG modulated current (mA) will influence image quality and to decide whether the electrocardiograph (ECG) modulated mA will effectively reduce the exposure dosage. Methods The cardiac pulsating phantom was set at three speed levels, i.e. high, medium, and low speed so as to simulate different heart rates. The phantom was scanned with ECG modulated mA turned on and off, and the exposure dosage of each scan sequence was documented. The images were reconstructed with reconstruction algorithm that matched the different levels of heart rate. CT values and their corresponding standard deviations at uniform areas on the images and the variation of the CT values at different locations were measured. The results from the two groups with and without ECG modulated mA were analyzed. Results Under the same level of heart rate, the exposure dosage was remarkably reduced when the ECG modulated mA was on than when it was off. Statistical analysis showed no significant difference (P 〉 0. 05) between the images from the two groups. Conclusion When scanning the heart with volume CT ( VCT), the application of ECG modulated mA can effectively reduce the exposure dosage without sacrificing the image quality.
出处 《中华放射学杂志》 CAS CSCD 北大核心 2006年第9期974-976,共3页 Chinese Journal of Radiology
关键词 心脏 体模 显像术 心电描记术 体层摄影术 X线计算机 Heart Phantoms, imaging Electrocardiograph Tomography, X-ray computed
  • 相关文献

参考文献4

二级参考文献11

  • 1Shrimpton PC and Edyvean S,CT scanner dosime try, Report on a round table meeting establishing a survey of standard dosimetic measurements on CT scanners,held at St George's Hospital,London,18 April 1997[J]. The British Journal of Radiology, 1998,71:1- 3.
  • 2Mahesh M, Scatarige JC, Cooper J, Fishman EK. Dose and pitch relationship for a particular multislice CT scanner[J]. AJR Am J Roentgenol 2001, 177: 1273- 1275.
  • 3Dawson P,Lees WR. Multi- slice Technology in Computed Tomography. Clinical Radiology 2001, 56: 302- 309.
  • 4Hui Hu.Multi- slice helical CT:Scan and reconstruction[J]. Medical Physics,1999,26( 1): 5- 18.
  • 5McNitt- Gray MF.AAPM/RSNA Physics Tutorial for Residents: Topics in CT: Radiation Dose in CT[J].RadioGraphics,2002, 22 (6): 1541 - 1553.
  • 6http://rsna2003.rsna.org/rsna2003/VBK/ conference/event- display.cfm?id=66601& p- navID=272& em_id=3106303.
  • 7朱昌寿.中国人受电离辐射照射剂量份额研究[J].中华放射医学与防护杂志,1998,18(5):340-345. 被引量:18
  • 8童新,孙志贤.辐射所致程序性细胞死亡的机制[J].国外医学(分子生物学分册),1995,17(3):131-135. 被引量:10
  • 9王晓峰,吴毅,杜国生.北京地区CT机应用质量检测与分析[J].中国辐射卫生,2001,10(2):112-114. 被引量:6
  • 10聂永康,蔡祖龙,赵绍宏.低剂量螺旋CT筛选检查早期肺癌的初步研究[J].中华放射学杂志,2002,36(3):230-234. 被引量:121

共引文献25

同被引文献357

引证文献45

二级引证文献385

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部