摘要
目的:比较高效液相色谱(HPLC)法和紫外分光光度计(UV)法测定孕酮阴道环体外释药量。方法:①HPLC法:十八烷基键合硅胶为填充剂,以甲醇∶水(63∶37)为流动相,流速为1.0ml/min,检测波长254nm;②UV法:检测波长254nm。结果:UV法测定释药量均值为9.9μg/ml,相对标准偏差(RSD)4.82%;HPLC法测定释药量均值为10.1μg/ml,RSD2.25%。配对t检验统计结果表明HPLC和UV法的结果没有显著性差异。结论:两种方法均可作为体外释放量的测试方法。
Objective: To compare HPLC with UV to determined the release quantity of progesterone vagina ring in vitro. Methods: 1) HPLC: An ODS column and a mobile phase of methanol-water (63 : 37) were used. The flow rate was 1.0 ml/min and measuring wavelength was 254 nm; 2) UV:The measuring wavelength was 254 nm. Results: The mean of release quantity was 10.1 μg/ml, RSD was 2.25% by HPLC and the mean of release quantity was 9.9 μg/ml, RSD was 4.82% by UV. Conclusion: There is no remarkable difference between the results of HPLC and UV according statistic conclusion with partnership t test.Two methods can both be used to measure the release quantity of progesterone vagina ring in vitro.
出处
《生殖与避孕》
CAS
CSCD
北大核心
2006年第9期546-549,553,共5页
Reproduction and Contraception
基金
国家十五科技攻关项目项目号:2004BA709B10
关键词
HPLC
UV
孕酮阴道环
体外释药量
HPLC
UV
progesterone vagina ring
release quantity in vitro