不同温湿度环境中布类包装压力蒸汽灭菌物品有效期的研究 被引量:6

Study on the Period of Validity of Autoclave Sterilization to Textile Packaged Medical Devices under Different Conditions of Temperatures and Humidity
下载PDF
导出
摘要 目的研究不同温度、湿度环境下,存放布类包装医用压力蒸汽灭菌物品的有效期,减少物品器材的损害,延长使用寿命,节省人力和资源。方法选择双层平纹纯棉外包布、内包布,包装医用手术器材包共440个。经预真空压力蒸汽灭菌器灭菌后,放置在各科室Ⅱ类、Ⅲ类环境的无菌物品柜内,每天测量存放环境的温度、湿度,保存15d、16d、17d、18d、19d、20d分别采样,进行实验室细菌培养,以确定灭菌物品使用的有效期限。结果440个灭菌包保存15~20d,细菌培养均为阴性。结论布类包装压力蒸汽灭菌物品,存放在温度为13~32℃,相对湿度为44%~93%的环境中,使用期限可以延长至20d。 Objective To study the period of validity of autoclave sterilization to textile packaged medical devices under different conditions of temperatures and humidity so as to decrease devices damage, lengthen their lifespan and save the manpower and the resources. Methods 440 double tabby-cotton packing sheets, inner packing sheets and packed medical surgery equipments were chosen. After sterilized by vacuum high pressure steam sterilizer, they were deposited in the aseptic goods cabinet of type Ⅱ and type III. Environment temperature and humidity were measured every day. They were sampled on 15 d, 16 d, 17 d, 18 d, 19 d and 20 d after depositing, and laboratory bacteria culture was carried out in order to ascertain the validity of aseptic goods. Results Bacteria culture of 440 antiseptic packages were negative from 15 -20 d after depositing. Conclusion The period of validity of autoclave sterilization of textile packaged medical devices may lengthen the use lifespan to 20 days under the environment of the temperature of 13-32℃ and the humidity of 44%-93%.
出处 《护理学报》 2006年第9期7-8,共2页 Journal of Nursing(China)
基金 深圳市龙岗区科技局2005年度科技计划项目(05-10)
关键词 压力蒸汽灭菌 布类包装 医疗器械 温度 湿度 有效期 autoclave sterilization textile packaged medical equipment temperature humidity validity
  • 相关文献

参考文献3

二级参考文献7

共引文献27

同被引文献32

引证文献6

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部