期刊文献+

利用阻抗/材料分析仪测量磁性薄膜的复磁导率 被引量:3

Measurement for complex permeability of magnetic thin-film by impedance/material analyzer
下载PDF
导出
摘要 该文提出了一种新的高频复磁导率的测试方法.首先确定E4991A射频阻抗/材料分析仪的测量原理为单线圈法,在此基础上推导出了测量有衬底磁性薄膜复磁导率所适用的公式.在实际测试过程中,将衬底放入夹具中作夹具短路补偿,减小了由衬底带来的误差;通过反复验证,弄清了系统提供的阻抗值表示的意义,并据此对磁导率的计算公式进行了修正,最后给出了磁性薄膜复磁导率的测量结果.实验表明,该方法简便易行,所得数据与文献中同类薄膜的测量数据基本符合. Abstract: This paper described a novel method for high - frequency complex permeability measurement. First, the theory for Agilent FA991A RF Impedance/Material Analyzer measurement, one -turn coil method, was studied. On the bases of this, formula of magnetic thin- film with substrate is derived for complex permeability calculation. At operating procedure, we put the substrate in the jig and did short compensation, decreasing the error introduced by it. We also found out the meaning of the impedance the system provided after proving. Formula for permeability calculation was thus modified. In the end, measurement result was presented. Operation of the method was easy, and the result was in coherence with that of similar thin - film from relevant papers.
出处 《实验技术与管理》 CAS 2006年第9期29-31,共3页 Experimental Technology and Management
关键词 复磁导率 磁性薄膜 射频 阻抗/材料分析仪 complex permeability magnetic thin film radio frequency Impedance/Material Analyzer
  • 相关文献

参考文献5

  • 1Fessant A,Gieraltowski J,Loaec J,et al.Influence of in-plane anisotropy and eddy currents on the frequency spectra of the complex perme ability of amorphous CoZr thin films[J].IEEE Trans Magn,1993,29(1):82-87.
  • 2Queffelec P,Gelin P,Gieraltowski J,et al.A microstrip device for the broad band simultaneous measurement of complex permeability and permittivity[J].IEEE Trans Magn,1994,30(2):224-231.
  • 3Huang H,Free C.E,G.Pitt K.E,et al.Relative permittivity measurements of thickfilm dielectrics at microwave frequencies[J].Electronics letters,1995,31(21):1812-1814.
  • 4张秀成,聂彦,何华辉,江建军.薄膜材料复介电常数与复磁导率测试研究[J].华中科技大学学报(自然科学版),2004,32(4):90-92. 被引量:12
  • 5M Yamaguchi,Y Miyazawa,K Kaminishi,et al.Soft magnetic applications in the RF range[J].Journal of Magnetism and Magnetic materials,2004,268:170-177.

二级参考文献6

  • 1Fessant A, Gieraltowski J, Loaec J. Influence of in plane anisotropy and eddy currents on the frequency spectra of the complex permeability of amorphous CoZr thin films. IEEE Trans Magn, 1993, 29(1): 82-87.
  • 2Laumo N J, Makovicka T J. Measurement of the permeability of thin film. Review of Scientific Instruments, 1999, 70(4) : 2 072-2 073.
  • 3Daiqing L, Charles F. A simple method for accurate loss tangent measurement of dielectrics using a microwave resonant cavity. IEEE Microwave and Wireless Components Letters, 2001, 11(3): 118-120.
  • 4Yabukami S, Uo T, Yarnaguchi M. High sensitivity permeability measurement of striped films obtained by input impedance. IEEE Trans Magn, 2001, 37(4) :2 776-2 778.
  • 5Yamaguchi M, Acher O, Miyazawa Y. Cross measure ment of thin film permeability Up to UHF range. J Magn Magn Mat, 2002, 242-245:970-972.
  • 6Ida N. Engineering electroma~gnetics. New York: Springer Verlag Inc., 2000.

共引文献11

同被引文献20

  • 1郑金菊,胡娟梅,余水宝,方允樟.提高铁基纳米晶高频磁导率的新方法[J].仪器仪表学报,2006,27(z2):1116-1117. 被引量:3
  • 2祁欣,刘明翠.巴克豪森效应和磁滞曲线的测定[J].实验技术与管理,2007,24(6):25-27. 被引量:8
  • 3Quefelec E Gelin P. A microstrip device for the broad band simultaneous measurement of complex permeability and permittivity [J]. IEEE Trans Magn, 1994, 30(2): 224--231.
  • 4Huang H, Free C E, Pitt K E, et al. Relative permittivity measurements of thick film dielectrics at microwave frequencies [J]. Electron Lett, 1995, 31(21): 1812--1814.
  • 5Fessant A, Gieraltowski J, Loaec J, et al. Influence of in-plane anisotropy and eddy currents on the frequency spectra of the complex permeability of amorphous CoZr thin films [J]. IEEE Trans Magn. 1993, 29(1): 82--87.
  • 6Lopes ME, Le Corvoisier P, Tabet JY,et al. Aldosterone and its antagonists in heart failure. Presse Med, 2003,32(2):79
  • 7Soberman J, Chafin CC, Weber KT. Aldosterone antagonists in congestive heart failure. Curr Opin Investig Drugs,2002,3(7):1 024
  • 8Stevenson L W, Perloff J K. The limited reliability of physical signs for the estimating hemodynamics in chronic heart failure. JAMA, 1989,261(6):884
  • 9王立锦.薄膜材料磁电阻特性测试系统的研制[J].实验技术与管理,2009,26(4):225-228.
  • 10Hurt D,Li S,Amann A.Versatile SQUID susceptometer with multiple measurement modes[J].IEEE TRANSACTIONS ON MAGNETICS,2013,49(7):3541-3544.

引证文献3

二级引证文献5

相关作者