期刊文献+

基于近红外三波长检测葡萄糖水溶液浓度的两种预测模型比较

The Comparison of Two Prediction Models Based on Near-Infrared Three Wavelengths of Glucose Concentrations Measurement in Aqueous Glucose Solutions
下载PDF
导出
摘要 讨论了近红外三波长测定水溶液中葡萄糖浓度的建模方法,采用偏最小二乘回归建立预测模型,对预测集样品中的葡萄糖浓度进行预测,结果所得预测均方根误差(RMSEP)为17.08mg/dL,浓度预测值与参考值的相关系数为0.997,而采用吸光度差值比率法所得预测模型的RMSEP达到54.94mg/dL,相应的相关系数仅为0.949。结果表明偏最小二乘回归所建模型用于近红外三波长葡萄糖浓度的预测,所得预测误差较小,初步验证了系统的可行性。 The modeling methods was discussed to determine glucose concentrations based on near-infared three wavelengths in aqueous glucose solutions in the text. The model was established based on partial least-square algorithm. When it was used to predict the glucose contents of validation set, the root mean square error of prediction (RMSEP) equaled 17.08mg/dL and the correlation coefficient between prediction values and reference values attained 0.997 ; while using logarithmic ratio of absorbency difference method to set up prediction model, the RMSEP reached 54.94mg/dL and the correlation coefficient equaled 0.949 merely. The results suggest that the model based on PLS algorithm has lesser RMSEP, it also validates the system's feasibility primarily.
出处 《生物医学工程研究》 2006年第2期85-88,共4页 Journal Of Biomedical Engineering Research
基金 国家自然科学基金资助项目(90508003) 教育部科学技术研究重点项目(02173)
关键词 近红外三波长 偏最小二乘 交互验证 葡萄糖浓度 Three-wavelengths in near-infrared Partial least-squares Cross-Validation Glucose Contents
  • 相关文献

参考文献10

二级参考文献18

  • 1许禄.化学计量学方法[M].北京:科学出版社,1997.101.
  • 2林方 丁海曙.-[J].世界医疗仪器,1998,4(4):20-20.
  • 3Pan S, Chung H, Arnold M A. Anal. Chem., 1996, 68: 1124.
  • 4Stephen F Malin, Timothy L Ruchti, Thomas B Blank, Suresh N Thennadil, Stephen L Monfre. Clinical Chemistry, 1999, 45(9): 651.
  • 5Arnold M A, Burmeister J, Small G W. Anal. Chem., 1998, 70: 1773.
  • 6Heise H M, Marbach R. Proc. of the Society of Photo-Optical Instrumentation Engineers, (SPIE), 1994, 2089: 114.
  • 7Heise H M, Marbach R, Koschinsky T H, Gries F A. Artif. Org., 1994, 18: 439.
  • 8Donald A Burns, Emil W Ciurczak. Handbook of Near-Infrared Analysis. New York: Marcfl Dfkker, Inc, 2001: 499.
  • 9Mark A. Arnold, Gary W.Small. Determination of Physiological Levels of Glucose in an Aqueous Matrix with Digitally Filtered Fourier Transform,Near-Infrared Spectra[J].Analytical Chemistry, 1990,62(14):1457-1464.
  • 10Kevin H.Hazen,Mark A.Arnold,Gary W.Small.Temperature-Insensitive Near-Infrared Spectroscopic Measurement of Glucose in Aqueous Solutions[J].Applied Spectroscopy,1994,48(4):477-483.