摘要
文章总结了上海SSDL自1983年至1994年期间参加国际原子能机构组织的各种剂量比对活动的目的、方法和结果.十多年来的比对数据表明,上海SSDL在放射治疗水平、辐射加工水平和个人剂量防护水平及对X、γ射线吸收剂量测量和控制等方面以满意的量值准确度与国际标准量值相统一。文章对测量误差进行了讨论.
The aims,methods and results of Shanghai SSDL participating inthe international dose intercomparisons, which were organized by IAEA in1983-1994, are reviewed .The data of dose intercomparison show that in therecent ten years,the values of dose determinated by Shanghai SSDL are in accordance with the international standard values on X and Y ray dosimetry with satisfied accuracy in the fields of radiotherapy,radiation procesaing and personal dose protection levels. Measurement errors are alsod iscussed
出处
《现代计量测试》
1996年第5期40-44,共5页
Modern Measurement and Test
关键词
辐射计量
剂量比对
测量
SSDL (Secondary Standard Dosimetry Laboratory),intercomparison, dosimetry, high-dose, TLD (Thermoluminescence Dosimeter),personal dose monitor, alanine/ESR dosimeter, IDAS (International Dose
Assurance Service)