期刊文献+

基于分段式变步长截断误差LMS算法的心率信号提取 被引量:1

The Abstraction of the Heart Rate Signals Based on Clipped-error LMS Algorithm with Variable Step Size of Segmenting Type
下载PDF
导出
摘要 自适应滤波对于心率信号这类非平稳随机信号处理是一种有效的方法。提出了一种分段式变步长截断误差LM S算法,采用自回归预测滤波器估计出通过超声多普勒传感器采集到的心率信号中的噪声,用它作为自适应噪声抵消系统中的参考信号,其结果是抵消了主输入通道中的部分噪声,使信噪比提高了36个dB左右,实现了心率信号的提取。 The adaptive filter is an effective method to dealing with such kind of non-steady random signal of the heart rate. The algorithm of clipped- error LMS algorithm with variable step size of segmenting type was proposed. We adopted auto-regression predict filter estimating the noise of heart rate signals through Doppler ultrasound sensor. We used it as the reference signal of the adaptive noise canceling system, its result is to remove some noise of main input, it made SNR raise by about 36 dB,It realized the abstraction of the heart rate signals.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2006年第5期974-976,共3页 Journal of Biomedical Engineering
关键词 心率信号 自适应噪声抵消 分段式变步长截断误差LMS算法 信噪比 收敛速度 Heart rate signal Adaptive noise canceling Clipped-error LMS algorithm with variable step size of segmenting type SNR Convergence rate
  • 相关文献

参考文献6

二级参考文献18

  • 1王布宏,郭英.频域LMS算法在语音消噪中的应用[J].空军工程大学学报(自然科学版),2000,1(3):64-67. 被引量:7
  • 2吴光弼,祝琳瑜.一种变步长LMS自适应滤波算法[J].电子学报,1994,22(1):55-61. 被引量:35
  • 3叶华,吴伯修.变步长自适应滤波算法的研究[J].电子学报,1990,18(4):63-70. 被引量:73
  • 4罗发龙 李衍达.神经网络信号处理[M].北京:电子工业出版社,1994..
  • 5Raymond Kwong H, Edward Johnston W. A variable step size LMS algorithm[J]. IEEE Trans." Signal Processing,1992, SP-40 (7): 1633-1642.
  • 6Haykin S, et al.. Adaptive filtering theory[M]. 1985.
  • 7Kalouptsidis N, Theodoridis S. Adaptive system identification and signal processing algorithms[M].Englewood Cliffs:Prentice-Hall, 1993.
  • 8Simon Haykin. Adaptive filter theory [M]. Englewood Cliffs:Prentice-Hall, 1996.
  • 9Chao J,Perez H, Tsujii S. A fast adaptive filter algorithm using eigenvalue reciprocals as step sizes[J]. IEEE Tram Acoust Speech Signal Process,1990,ASSP-38:1343-1352.
  • 10Chazan D, Medan Y,Shvadron U. Noise cancellation for hearing aids[J]. IEEE Tran Acoust Speech Signal Process, 1988,ASSP-36:1 697-1 705.

共引文献80

同被引文献11

  • 1王鹏,魏守水,黄青华.基于小波变换的自适应滤波器消除脉搏波基线漂移[J].中国医学物理学杂志,2004,21(5):296-299. 被引量:22
  • 2HYONYOUNG H, JUNG K. Artifacts in wearable photoplethysm- ographs during daily life motions and their reduction with least mean square based active noise cancellation method [J].Comput Biol Med, 2012, 42(4): 387-393.
  • 3KIM B S, YO S K. Motion artifact reduction in photoplethysmography using independent component analysis[J]. IEEE Trans Biomed Eng, 2006, 53(3): 566-568.
  • 4REDDY K A, KUMAR V J. Motion artifact reduction in photoplethysmographic signals using singular value decomposition [ C ]. Warsaw: IEEE Instrumentation and Measurement Technology, 2007: 1-4.
  • 5CHAN K W, ZHANG Y T. Adaptive reduction of motion artifact from photoplethysmographic recordings using a variable step-size LMS filter[C]. Florida: IEEE Sensors, 2002: 1343-1346.
  • 6RAM M R, MADHAV K V, KRISHNA E H, et al. A novel approach for motion artifact reduction in PPG signals based on AS- LMS adaptive filter [J]. IEEE Trans Instrum Meas, 2012, 61(5): 1445-1457.
  • 7YOUSEFI R, NOURANI M, PANAHI I. Adaptive cancellation of motion artifact in wearable biosensors [ C]. San Diego: 34 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2012: 2004-2008.
  • 8曹玉珍,蔡伟超,程旸.基于MEMS加速度传感器的人体姿态检测技术[J].纳米技术与精密工程,2010,8(1):37-41. 被引量:68
  • 9龚渝顺,吴宝明,高丹丹,蒋洪,闫庆广.动态环境血氧饱和度监测的运动干扰分离自适应对消方法[J].航天医学与医学工程,2012,25(4):266-270. 被引量:3
  • 10张会芝,穆晓霞,王鲜芳.一种新变步长LMS算法及在自适应波束形成中的应用[J].测控技术,2014,33(1):129-132. 被引量:5