期刊文献+

汉防己甲素选择性PVC膜电极的研制

Preparation of PVC Membrane Tetrandrine Ion Selective Electrode
下载PDF
导出
摘要 以四苯硼酸-汉防己甲素离子缔合物作活性材料,制成了PVC膜汉防己甲素电极。对影响电极性能的诸因素进行了讨论,以最佳条件制备的电极其线性区为5.0×10-3~5.0×10-6mol/L,平均斜率为26.5mV/pC,检测下限为1.0×10-7mol/L,在pH1.0~5.5范围内适用。电极抗干扰能力强,响应快,重现性和稳定性较好,适用于汉防己甲素制剂的快速测定。 Using tetrandrine tetraphenylborate ion pair complex as active material,PVC membrane tetrandrine ion selective electrode was prepared. The factors associating with the electrode response were discussed in detail.The electrode shows Nerstian response over the tetradrine concentration range from 5.0×10 -3 to 5.0×10 -6 mol/L in the pH range of 1.0 to 5.5,with a slope of 26. 5mV/decade, the limit of detection being 1.0×10 -7 mol/L.Regular cations and organic acids show little interference. The electrode gives fast response and good reproducibility,and has long life span. It is suitable for rapid determination of tetrandrine samples.
出处 《广东药学院学报》 CAS 1996年第4期219-221,共3页 Academic Journal of Guangdong College of Pharmacy
关键词 汉防已甲素 PVC膜电极 电极 研制 tetrandrine tetrandrine-tetraphenylborate ion selective electrode