不同频率和强度超声波溶栓效果的体外实验 被引量:3

Sonothrombolysis at different frequencies and intensities: an experimental study in vitro
下载PDF
导出
摘要 目的探讨体外实验条件下超声频率和强度对其溶栓效率的影响。方法取健康人全血标本56份,37℃恒温水浴孵育2h后形成体外血栓。样本被分为5个辐照组和1个对照组。辐照组分别以0.7W/cm2,0.5MHz;0.7W/cm2,1MHz;0.7W/cm2,2MHz;1.4W/cm2,2MHz和1.8W/cm2,2MHz的脉冲式超声波照射10min;对照组无超声处理。计算血栓溶栓率,比较各相同声强不同频率组之间差异和相同频率不同声强组之间差异,并分别测溶栓率与超声波频率及强度之间的关系。结果声强0.7W/cm2条件下,频率为0.5MHz、1MHz和2MHz的超声波有明确的溶栓效果,溶栓率与频率呈负相关(r1=1.000,P<0.01);频率2MHz条件下,声强为0.7W/cm2、1.4W/cm2和1.8W/cm2的超声波有明确效果,溶栓率与声强呈正相关(r2=0.980,P<0.05)。结论在一定条件下,超声波有明确的直接溶栓效果,并与频率和声强相关。诊断用超声波(频率2MHz、声强<2W/cm2)可能具有潜在的溶栓价值。 Objective To evaluate the effects of frequency and intensity of ultrasound on sonothrombolysis in vitro. Methods Fifty - six thrombi samples of whole blood from the healthy volunteers were formed by being incubated in 37℃ water bath for 2 hours. The samples were 5 irradiation groups and a control. The irradiation groups were respectively radiated by the pulse ultrasound at 0.7 W/cm^2 - 0.5 MHz, 0.7 W/cm^2 - 1 MHz, 0.7 W/cm^2 - 2 MHz, 1.4 W/cm^2 - 2 MHz and 1.8 W/cm^2 - 2 MHz for 10 minutes and the control was treated without ultrasound. The thrombolysis rates were caculated and compared between the groups. And the correlation between the thrombolysis rate and the frequency and the intensity of ultrasound were calculated by SPSS 10.0 software respectively. Results When the intensity was kept at 0.7 W/cm^2 , ultrasound of 0.5 MHz, 1 MHz and 2 MHz solved the thrombi obviously and there was a negative correlation between the thrombolysis rate and frequency ( r1 = 1.0(30, P 〈 0.01 ). And when the frequency was kept at 2 MHz, ultrasound of 0.7 W/cm^2, 1.4 W/cm^2 and 1.8 W/cm^2 also solved the thrombi significantly and there was a positive correlation between the thrombolysis rate and the intensity ( r2 = 0. 980, P 〈 0.05 ). Conclusion Ultrasound can lysis thrombi directively in certain conditions and its effect is related with the frequency and the intensity. Diagnostic ultrasound (2 MHz, 〈 2 W/cm^2 ) has potential to be applied in the treatment to thrombi.
出处 《临床超声医学杂志》 2006年第11期641-643,共3页 Journal of Clinical Ultrasound in Medicine
基金 国家自然科学基金(30500478)
关键词 血栓溶解疗法 声强 频率 溶栓率 Thrombolytic Intensity Frequency Thrombolysis rate
  • 相关文献

参考文献8

  • 1Behrens S, Spengos K, Daffertshofer M, et al. Trartscranial ultrasotmd-improved thrombolysis: diagnostic vs. therapeutic uhrasound. Ultrasound Med Biol, 2001, 27(12): 1683-1689.
  • 2Rosenschein U, Furman V, Kemer E, et al. Ultrasound imaging-guided noninvasive ultrasotmd thrombolysis. Preclinical results. Circulation, 2000,102:238 - 245.
  • 3Porter TR, Kricsfield D, Li S, et al. Comparison of d-dimer activity and residual particle size produced following thrombus disruption by perfluorocarbon containing microbubbles versus urokinase in the presence of low frequency ultrasound. J Am Coll Cardiol, 1998, 31:219.
  • 4Suchkova V, Carstensen E L, Francis C W. Ultrasound enhancement of fibrinolysis at frequencies of 27 to 100 kHz. Ultrasound Med Biol, 2002, 28(3) : 377 - 382.
  • 5Daffertshofer M, Hennerici M. Ultrasound in the treatment of ischemic stroke. Lancet Neurol, 2003, 2(5): 283 - 290.
  • 6华兴,高云华.超声波与声学造影剂在溶栓中的作用[J].临床超声医学杂志,2006,8(7):425-427. 被引量:9
  • 7Mexandrov. Ultrasound-enhanced thrombolysis for stroke: clinical significance. Euro J Ultrasound, 2002, 16:131 - 140.
  • 8夏红梅,高云华,卞爱娜,刘平,谭开彬,刘政.亲血栓性靶向超声造影剂增强兔腹主动脉新鲜血栓显像的实验研究[J].中国超声医学杂志,2004,20(11):810-813. 被引量:11

二级参考文献32

  • 1夏红梅,高云华,卞爱娜,刘平,谭开彬,刘政.亲血栓性靶向超声造影剂增强兔腹主动脉新鲜血栓显像的实验研究[J].中国超声医学杂志,2004,20(11):810-813. 被引量:11
  • 2于锐,程远.超声和超声微泡造影剂的溶栓作用[J].临床超声医学杂志,2004,6(6):363-365. 被引量:7
  • 3高云华,谭开彬.超声造影剂介导靶向治疗的研究进展[J].中华超声影像学杂志,2004,13(12):951-952. 被引量:2
  • 4Edler I.Echocardiography,a historical perspective.In:Schapira JN,Harold JG,editors.Two-dimentional Echocardiography and Cardiac Doppler.Baltimore:Williams and Willkins,1989,1 ~ 33.
  • 5Trubestein G,Engele C,Etzel F,et al.Thrombosis by ultrasound.Clin Sci Mol Med,1976,51:697~698.
  • 6Porder JB,Porder KN,Meltzer RS.Ultrasound bioeffects.Echocardiography,1987,4:89~99.
  • 7Rosenschein U,Furman V,Kerner E,et al.Ultrasound imaging-guided noninvasive ultrasound thrombolysis.Preclinical results.Circulation,2000,102:238 ~ 245.
  • 8Porter TR,Kricsfield D,Li S,et al.Comparison of d-dimer activity and residual particle size produced following thrombus disruption by perfluorocarbon containing microbubbles versus urokinase in the presence of low frequency ultrasound.J Am Coll Cardiol,1998,31:219.
  • 9Simon S,Stefan K,Lutz K,et al.Influence of ultrasound operating parameters on ultrasound-induced thrombosis in vitro.Ultrasound Med Biol,2005,6(31):841~847.
  • 10Takahama N,Hirai M,Hayashi H.A basic experiment of thrombolytic therapy by ultrasound.Jpn Circ J,1989,53:182.

共引文献16

同被引文献18

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部