期刊文献+

Activity optimization method in SPECT:A comparison with ROC analysis

Activity optimization method in SPECT:A comparison with ROC analysis
下载PDF
导出
摘要 A discriminant method for optimizing activity in nuclear medicine studies is validated by comparison with ROC (received operating characteristic)-curves. The method is tested in 21 single photon emission computerized tomography (SPECT), performed with a cardiac phantom. Three different lesions (Lb L2 and L3) were placed in the myocardium-wall by pairs fbr each SPECT. Three activities (84, 37 or 18.5 MBq) of 99mTc were used as background. Linear discriminant analysis was used to select the parameters that characterize image quality among the measured variables in the images [(Background-to-Lesion (B/Li) and Signal-to-Noise (S/N) ratios)]. Two clusters with different image quality (P=0.021 ) were obtained. The ratios B/Lj, B/L2 and B/L3 are the parameters used to construct the function with 100% of cases correctly classified into the clusters. The value of 37 MBq was the lowest tested activity for which good results for the B/Li ratios were obtained. The result coincides with the applied ROC-analysis (r=0.89). A discriminant method for optimizing activity in nuclear medicine studies is validated by comparison with ROC (received operating characteristic)-curves. The method is tested in 21 single photon emission computerized tomography (SPECT), performed with a cardiac phantom. Three different lesions (L1, L2 and L3) were placed in the myocardium-wall by pairs for each SPECT. Three activities (84, 37 or 18.5 MBq) of 99mTc were used as background. Linear discriminant analysis was used to select the parameters that characterize image quality among the measured variables in the images [(Background-to-Lesion (B/Li) and Signal-to-Noise (Si/N) ratios)]. Two clusters with different image quality (P=0.021) were obtained. The ratios B/L1, B/L2 and B/L3 are the parameters used to construct the function with 100% of cases correctly classified into the clusters. The value of 37 MBq was the lowest tested activity for which good results for the B/Li ratios were obtained. The result coincides with the applied ROC-analysis (r=0.89).
出处 《Journal of Zhejiang University-Science B(Biomedicine & Biotechnology)》 SCIE CAS CSCD 2006年第12期947-956,共10页 浙江大学学报(英文版)B辑(生物医学与生物技术)
基金 Project supported by the Third World Academy of Sciences (TWAS), Cuba
关键词 Activity optimization Image quality SPECT Discriminant analysis ROC curves 活度优化 图像质量 SPECT 判别式分析 ROC曲线
  • 相关文献

参考文献10

  • 1Venables, W,Ripley, B.Statistics and Computing. Modern Applied Statistics with S-Plus[]..1994
  • 2Moonen, M,Jacobsson, L.Effect of administered activity on precision in the assessment of renal function using gamma camera renography[].Nucl Med Commun.1997
  • 3Metz,C.Basic principles of ROC-analysis[].Sem Nucl Med.1978
  • 4Juez, P,Díaz, F.Probabilities and Statistic in Medicine Applications in the Clinic Practice and Sanitary[].Madrid.1996
  • 5NEMA (National Electrical Manufacturers Association).Performance Measurements of Scintillation Cameras[]..1994
  • 6Stark, S,Carlsson, S.Digital filtering of bone scans: a ROC study[].Nucl Med Commun.1997
  • 7Tsui, B,Zhao, X.Quantitative single-photon computed tomography: basics and clinical considerations[].Sem Nucl Med.1994
  • 8Hanley, J,McNeil, B.The meaning and use of the area under a receiver operating characteristic (ROC) curve[].Radiology.1982
  • 9Mattsson, S,Jacobsson, L,Vestergren, E.The basic principles in assessment and selection of reference doses: considerations in nuclear medicine[].Rad Prot Dosim.1998
  • 10Perez, M,Quevedo, J,Diaz-Rizo, O,Dopico, R,Estévez, A,Viamonte, A.Administered activity optimization in skeletal scanning using MDP labelled 99m-Tc[].ALASBIMN J.2002