摘要 应用循环伏安法(CV)和示差脉冲吸附阴极溶出伏安法(DPAdCSV),研究头孢唑啉钠在悬铜汞齐微电极(HCADME)上的电化学行为及电极反应机理,发现头孢唑啉钠与Cu(Ⅱ)形成配比为1∶1的配合物,并吸附在HCADME上,反应电子数为2。并建立了示差脉冲吸附阴极溶出伏安法(DPAdCSV)测定痕量头孢唑啉钠的灵敏方法。其测定的线性范围为6×10 ̄(-1)~1×10 ̄(-9)mol/L;检测限约为4×10 ̄(-10)mol/L(t_(ac)=120s)。本法用于注射剂中该物质的测定,结果满意。用本法直接测定尿中头孢唑啉钠是可能的。 The electrochemical behavior and the mechanism of the electrode reaction of cefazolin sodium were investigated at hanging copper amalgam dropping microelectrode(HCADME)in HCl-KCl(pH1.1)buffer solution. It was found that cefazolin sodium could form complex with Cu(Ⅱ)and be adsorbed at the electrode surface.The coordination number is 1 and the number of transfer electron involved in the reaction is 2.This methode was successfully applied to the determination of trace amounts of cefazolin sodium by differential pulse adsorption cathodic stripping voltammetry (DPAdCSV). The Iinear range determined is from 1× 10 ̄(-9)to 6×10 ̄(-5)mol/L;The detection limit is about 4×10 ̄-(10)mol/L(tac=120s). The method was applied to the determination of cefazolin sodium in injection with satisfactorv Yesults, It is possible that the direct determination of cefazolin sodium in mimic urine sampke was carried out with the recommended method.
出处 《南通医学院学报》 1996年第4期464-467,共4页 ACTA Academiae Medicinae Nantong
基金 国家自然科学和教委博士点基金
关键词 悬铜汞齐微电极 头孢唑啉钠 伏安法 药物分析 Hanging copper amalgam dropping microelectrode(HCADME) Cefazolin sodium Differential pulse adsorption cathodic stripping voltammetry(DPAdCSV)
  • 相关文献

参考文献1

二级参考文献8

  • 1庞志功,化学通报,1993年,9卷,39页
  • 2姚金玉,分析测试通报,1990年,3卷,9期,35页
  • 3甘知行,化学学报,1989年,47卷,78页
  • 4谷林瑛,电化学方法.原理及应用,1986年
  • 5陈洪渊,Anal Chem,1984年,319卷,240页
  • 6李南强,化学学报,1984年,42卷,1057页
  • 7陈洪渊,Anal Chem,1983年,314卷,657页
  • 8李南强,化学学报,1983年,41卷,351页

共引文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部