摘要
采用长1m,内径2mm,内装5%OV-107/GasChromQAWDMCS(150~180目)的玻璃柱,以氢火焰离子化检测器,在相同色谱条件下,对三唑酮和甲基异柳磷进行了定量分析,其标准偏差和变异系数分别为0.014、0.008和0.33%、0.77%,回收率分别为100.2%~101.8%和98.7%~101.3%
A method was described for the determination of triadimefon and isofenphosmethyl by gas chromatography with FID on 5% OV-107/Gas Chrom Q AW DMCS(150~180m)glassic colum.Dibutyl o phthalate was adopted as the internal standard.The standard deviation was 0.014,0.008 cespectively;and the coefficient of variation was 0.33% and 0.77% respectively.The average recovery was 101.0%and 100% respectively
出处
《安徽农业科学》
CAS
1996年第4期339-340,共2页
Journal of Anhui Agricultural Sciences