NICO无创心输出量监测系统在ICU中的应用研究 被引量:3

The application study of novametrix medical systems inc(NICO) in ICU
下载PDF
导出
摘要 目的研究NICO无创心输出量监测系统在ICU中的应用。方法重症监护病房(ICU)15例危重患者,在呼吸机辅助控制通气条件下,采用美国NICO心输出量监测系统,用部分二氧化碳重复吸入法(RB)连续监测无创心输出量(RBco),并同时采用温度稀释法测定心输出量(TDco),比较两种方法所测心输出量值的相关性。结果统计分析显示两种方法测定的心输出量值无显著性差异(P>0.05),相关回归分析显示相关系数为0.90,回归方程为RBco=0.81TDco+1.25,偏离为0.20,精确度为0.78。结论NICO监测系统所测无创心输出量(RBco)与温度稀释法心输出量(TDco)有良好的相关性,NICO无创心输出量监测系统为ICU提供了一种安全可靠的无创心输出量测定手段。 Objective To study the application of Novametrix Medical Systems Inc(NICO) in ICU .Methods Fifteen critical patients receiving mechanical ventilation were studied . Each patients was instrumented with a pulmonary artery thermodilution(TD) cathether(DualTherm, Braun Medical Inc, Bethlehem , PA ), and a partial CO2 rebreathing (RB) ventilator circuit(Novametrix Medical Systems, Wallingford, CT) . Cardiac output was measured simultaneous by RB and TD methods. Paired ttest and regression statistics were used to compare the degree of agreement between concurrent measurement of TDco and RBco. Results Paired t - test analysis revealed no significant difference between RBco and TDco measurement (P〉0.05). Regression analysis gives an r=0.90 with a slop of 0.81 and an intercept of 1.25L/min. Bland-Airman analysis resulted in a bias of 0.20 L/min with a precision (ISD) of 0.78L/min. Conclusion There was a high correlation between RBeo and TDco . Novametrix Medical Systems Ine(NICO)can provide a safe and reliable estimation of noninvasive cardiac output in ICU.
作者 刘琼 黄纪坚
出处 《医疗设备信息》 2006年第11期15-16,共2页 Information of Medical Equipment
关键词 NICO无创心输出量监测系统 心输出量 ICU novametrix medical systems inc(N ICO) cardiac output ICU
  • 相关文献

参考文献6

  • 1刘琼,黄纪坚.无创心输出量监测的进展与评价[J].中国呼吸与危重监护杂志,2003,2(6):379-381. 被引量:16
  • 2李涛.心电监护数据的分析[J].医疗设备信息,2004,19(10):10-12. 被引量:3
  • 3陈国珍,孙锟,陈树宝,张玉奇,洪雯静,薛海虹.三维超声估测心室容量与室壁质量的实验研究[J].医疗设备信息,2005,20(7):2-5. 被引量:2
  • 4Osterlund B,Gedeon A,Krill P,et al.A new method of using gas exchange measurement for the noninva sive determination of cardiac output:clinical experiences in adults following cardiac surgery.Acta Anaesthesiol Scand[J].1995,39:727-732.
  • 5Robert G Loeb,Elizabeth A Brown,James A DiNardo,et al.Clinical accuracy of a new non-invasive cardiac output monitor[J].Anesthesiology,1999,91(3A):474.
  • 6Monica Botero,Kurt Briesacher,David Kirby,et al.Comparison of noninvasive determination of cardiac output(CO) using a partial CO2 rebreathing system with thermodilution during cardiac surgery[J].Anesthesiology,2000,(2S):32-34.

二级参考文献35

  • 1陈国珍,孙锟,陈树宝,张玉奇,洪雯静,薛海红.心脏TTO三维超声成像影响因素的实验研究[J].医疗设备信息,2004,19(8):45-47. 被引量:2
  • 2胡大一,郭成军,李瑞杰.心率变异性测量标准,生理释义与临床应用(二)[J].中国医疗器械信息,1997,3(5):17-20. 被引量:5
  • 3孙锟,陈树宝.二维超声技术检测右心室容量[J].中华医学杂志,1993,73(9):535-540. 被引量:11
  • 4杨国兴.Visual C++ 6.0程序设计实训教程[M].北京:中国水利水电出版社,..
  • 5张红军.Visual C++编程详解 [M].北京:科学出版社,..
  • 6钱能.C++程序设计教程 [M].北京:清华大学,..
  • 7Lange A, Palka P, Burstow DJ,et al. Three-dimensional echocardiography: historical development and current applications[J]. J Am Soc Echocardiogr, 2001. 14(5): 403-412.
  • 8Jiang L, Vazquez de Prada JA, Handsehumaeher MD, et al, Three-dimensional echocardiography: in vivo validation for right ventrieular free wall mass as an index of hypertrophy[J]. J Am Coll Cardiol, 1994, 23(7) : 1715 - 1722.
  • 9Teupe C, Takeuehi M, Ram SP, et al. Three dimensional echocardiography : in - vitro validation of a new, voxel - based method for rapid quantification of ventrieular volume in normal and aneurysmal left ventricles[J]. Int J Card Imaging, 2001, 17(2):99 - 105.
  • 10Teupe C, Takeuchi M, Yao J, Pandian N. Determination of left ventrieular mass by three dimensional echocardiography: in vitro validation of a novel quantification method using multiple equi -angular rotational planes for rapid measurements[J]. Int J Card Imaging, 2002, 18(3): 161-167.

共引文献18

同被引文献11

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部