声通道上的肋骨对高强度聚焦超声在羊肝内形成凝固性坏死的影响 被引量:5

Effect of ribs in high intensity focused ultrasound beam path on formation of coagulative necrosis in goat liver
原文传递
导出
摘要 目的从有效性、安全性和剂量学的角度探讨声通道上的肋骨对高强度聚焦超声(HIFU)损伤山羊肝的影响。方法HIFU分别辐照声通道上有肋骨(对照组)和声通道上肋骨被手术切除(实验组)的羊肝。辐照结束后2d处死动物,观察靶区凝固性坏死形成情况,测量凝固性坏死的体积并计算能效因子(energyefficiencyfactor,EEF)。结果实验组的凝固性坏死形成率明显高于对照组,EEF远远小于对照组。结论声通道上的肋骨明显影响凝固性坏死的形成率和EEF,提高HIFU治疗受肋骨阻挡的肝癌的治疗效率非常重要。 Objective To explore the effect of ribs in high intensity focused ultrasound(HIFU) beam path on HIFU ablation of goat liver. Methods A model-JC focused ultrasound tumor therapeutic system was used. A 0.75 MHz focused transducer with 150mm aperture and 120ram focal length was used in all experiment. Acoustical power can be adjusted. Twenty goats were divided into control group(HIFU beam through rib cage) and experiment group(Ribs in HIFU beam path were surgically removed). Twenty targeted regions at 50mm away from skin surface were applied for creating necrosis with linear scanning of 15mm length using HIFU in two groups. All animals were sacrificed two days later and targeted organs were dissected. After obtaining the maximal section, the volumes of the necrotic regions were measured, then to calculate energy efficiency factor(EEF). Results For experiment group, the proportion of formed coagulative necrosis relative to number of targeted region to be treated was bigger than control group, and EEF was more smaller than control group. Conclusions Ribs in HIFU beam path affected markedly the formation of coagulative necrosis and enhanced EEF, that indicate that the treatment efficiency can be decreased when treating liver cancer shielded by ribs with HIFU.
出处 《中华超声影像学杂志》 CSCD 2006年第12期943-945,共3页 Chinese Journal of Ultrasonography
基金 国家杰出青年基金资助项目(30325027) 自然科学基金(30471653)
关键词 高强度聚集超声 肋骨 坏死 High intensity focused ultrasound Ribs Liver Necrosis
  • 相关文献

参考文献6

二级参考文献20

  • 1黄孝迈.手术学全集胸外科卷[M].北京:人民军医出版社,1995.224-254.
  • 2Wang Z,Bai J,Li F,et al.Study of a "biological focal region" of high-intensity focused ultrasound.Ultrasound in Medicine and Biology,2003,29(5):749-754.
  • 3Wei K,Skyba DM,Firschke C,et al.Interaction between microbubbles and ultrasound:in vitro and in vivo observation.J Am Coll Cardiol,1997,29:1081-1088.
  • 4Zha Daogang,Liu Yili,Chen Shuyuan,et al.Effect of ultrasound exposure on sonicated microbubbles stability.F Med Coll PLA,2000,15(4):247-249.
  • 5Zabari GA, Riche A, Zizzi HC, et al. Surgical and nonsurgical management of primary and metastatic liver tumor. Am Surg, 1998,64: 211 ~ 221.
  • 6Zibarl GB,Riche A,Zizzi HC,et al.Surgical and nonsurgical management of primary and metastatic liver tumors. Am Surg, 1998, 64:211-220.
  • 7程树群,周信达,汤钊猷,余耀,鲍苏苏,钱德初.碘化油协同高功率聚焦超声对肝癌的治疗作用[J].中华实验外科杂志,1998,15(1):23-24. 被引量:8
  • 8伍烽,陈文直,白晋,刘长安,罗亿治,邹建中,徐昆,朱辉,王芷龙,王智彪.高强度聚焦超声治疗原发性肝癌的初步临床研究[J].中华超声影像学杂志,1999,8(4):213-216. 被引量:153
  • 9白晋,伍烽,王智彪,阮祥燕,杜永洪,孔繁斌,王芷龙,陈迅,童艺,齐家俊,刘川,胡凯,邹建中,王文见,彭兴礼,陈文直,王志刚.高强度聚焦超声对香猪正常肝组织辐照时间与损伤范围关系的探讨[J].中华超声影像学杂志,1999,8(4):247-250. 被引量:8
  • 10孔凡斌,伍烽,王智彪.高强度聚焦超声治疗肝脏肿瘤[J].临床超声医学杂志,2001,3(1):53-55. 被引量:14

共引文献27

同被引文献31

  • 1秦海峰,黄乃祥,汪伟,罗明玉.肋骨和肺组织遮挡对高强度聚集超声生物学焦域的影响[J].中华实验外科杂志,2004,21(10):1251-1252. 被引量:1
  • 2李发琪,王智彪.高强度聚焦超声治疗关键技术的思考和完善[J].中国医学文摘(耳鼻咽喉科学),2006,21(2):91-93. 被引量:10
  • 3李国逊,叶启发,牛英,宫念樵,王西墨,金中奎,龙刚,朱劲钢,王家林,王宏磊,王萍,李刚,张宇,龚谨,陈实,江涛.同种体外猪肝脏灌流中对于体外肝脏热切取与冷切取效果的比较研究[J].中国现代医学杂志,2007,17(12):1446-1449. 被引量:2
  • 4Wu F, Wang Z B, Chen W Z,et a l.Extracorporeal high intensity focused ultrasound ablation in the treatment of patients with large hepatocellular carcinoma [J].Ann Surg Oncol, 2004,11 ( 12 ) : 1061-1069.
  • 5Pelletier G,Ducreux M,Gay F,et al. Treatment of unresectable hepatocellular carcinoma with lipiodol chemoembolization : a multicenter randomized trial [J]. Group C HC J Hepatol, 1998,29: 129-134.
  • 6Wu F,Wang Z B, Chen W Z,et al. Advanced hepatocellular carcinoma: treatment with high-Intensity focused ultrasound ablation combined with transcatheter arterial embolization[J]. Radiology, 2005,235 : 659-667
  • 7Wu F,Wang Z B,Chen W Z,et al. Extracoporeal high intensity focused ultrasound ablation in the treatment of 1038 patients with solid carcinomas in China: an overview[J]. Ultrasonics Sonochemistry,2004,11: 149-154.
  • 8Wu F, Chen W Z, Bai J, et al. Pathological changes in human malignant carcinoma treated with high intensity focused ultrasound[J]. Ultrasound Med Biol, 2001,27 : 1099.
  • 9ter Haar G.Acoustic surgery.Phys Today,2001,54:29-34.
  • 10Kennedy JE.High-intensity focused ultrasound in the treatment of solid tumors.Nat Rev Cancer,2005,5:321-327.

引证文献5

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部