摘要 利用正交试验设计测试WFX-120原子吸收分光光度计测定锰的操作参数。正交试验设计是研究和处理多因素试验的一种科学方法。采用L9(34)正交表得出最佳工作条件为灯电流1.9 mA、乙炔流量/空气流量1.8/6.0 L/min、狭缝宽度0.2 mm、燃烧器高度6 mm。 Orthogonal test is one of the scientific methods in studying and treating muhifaetor tests. Optimum operation conditions of determination the manganese ion by flame atomic absorption were researched by means of orthogonal test with four factors and four levels. In the orthogonal layout of L9 (3^4 ), the results show that optimum operation conditions of determination the manganese ion by flame atomic absorption on WFX - 120 atomic absorption spectrometer are that lamp current is 1.9 mA, width of slit is 0.2 mm, height of the burnet 5 nun and the ratio of flowing of acetylene to flow of air 1.8/6.0 L/min.
出处 《当代化工》 CAS 2006年第6期433-435,438,共4页 Contemporary Chemical Industry
关键词 原子吸收分光光度法 正交试验设计 flame atomic absorption spectrophotometry manganese ion orthogonal test design