期刊文献+

BODPOD体成分测量仪中两种方法测量成年人身体成分的差异性及相关性研究 被引量:3

Study on Difference and Correlation of Two Methods in Percent Fat Estimation on Adults in BODPOD
下载PDF
导出
摘要 目的:研究在BODPOD体成分测量仪中,利用估计测量与直接测量两种方法测得的成年人身体成分结果是否存在显著性差异,探索估计测量方法的准确性;方法:利用BODPOD体成分测量仪测量35名受试者的身体成分。分别采用估计测量和直接测量两种方法测得胸腔内气体体积,然后再分别计算出体脂率。比较估计测量和直接测量两种方法所得的体成分值的差异性与相关性;结果:两种方法所得的%BF值无显著性差异,而是存在高度相关性(r=0.954,P<0.001)。结论:估计测量方法也是一种较为精确的测量体成分的方法。从长远来看,存在估计测量方法代替直接测量方法测量体成分的可能。 The purpose of this study was to estimate the body fat percent (%BF) through using the predication and measurement thoracic gas volumes in BOD POD in order to find out that whether two methods have significant difference on adults in BODPOD. The body fat percent of thirty-five subjects were measured by using BODPOD. The datum of thoracic gas volume (Vtg) was estimated by using predict (Vtgpred) and measure (Vtgmeas) ,then calculated the value of %BF. Following that,two methods were compared and get the difference and correlation of the value of %BF. The result showed that there was no significant difference of subjects value of BF by using Vtgpred and Vtgmeas methods,while these values have significant correlation with the two methods (r = 0. 954,P〈0. 001). It concluded that Vtgpred method was also a precise method of measuring %BF. From the long views, it is possible to replace the Vtgmeas by Vtgpred.
出处 《体育科学》 CSSCI 北大核心 2006年第12期66-69,共4页 China Sport Science
基金 上海市科技启明星计划资助项目(03QD14046) 上海市重点学科建设项目资助(T0901)
关键词 体成分BODPOD 测量 方法 成年 body composition thoracic gas volume BODPOD
  • 相关文献

参考文献20

  • 1FIELDS D A,HULL H R,CHELINE A J,et al.Child-specific Thoracic Gas Volume Prediction Equations for Air-displacement Plethysmography[J].Obes Res,2004,12 (11):1797-1804.
  • 2COLLINS M A,MILLARD-STAFFORD M L,EVANS E M,et al.Effect of Race and Musculoskeletal Development on the Accuracy of Air Plethysmography[J].Med Sci Sports Exe,2004,36 (6):1070-1077.
  • 3FIELDS D A,HIGGINS P B,HUNTER G R.Assessment of Body Composition by Air-displacement Plethysmography:Influence of Body Temperature and Moisture[J].Dyn Med,2004,3 (1):3.
  • 4COLLINS A L,SAUNDERS S,MCCARTHY H D,et al.Within and Between Laboratory Precision in the Measurement of Body Volume Using Air Displacement Plethysmography and Its Effect on Body Composition Assessment[J].Int J Obes Relat Metab Disord,2004,28 (1):80-90.
  • 5BOSY-WESTPHAL A,MAST M,EICHHORN C,et al.Validation of Air-displacement Plethysmography for Estimation of Body Fat Mass in Healthy Elderly Subjects[J].Eur J Nutr,2003,42(4):207-216.
  • 6VESCOVI J D,ZIMMERMAN S L,MILLER W C,et al.Effects of Clothing on Accuracy and Reliability of Air Displacement Plethysmography[J].Med Sci Sports Exe,2002,34 (2):282-285.
  • 7MEGAN A,MCCRORY,PAUL A,et al.Body Composition by Air-displacement Plethysmography by Using Predicted and Measured Thoracic Gas Volumes[J].J Appl Phy,1998,84 (4):1475-1479.
  • 8李艳平,马冠生.BOD POD体成分测量仪的可靠性和有效性[J].国外医学(卫生学分册),2003,30(2):118-123. 被引量:3
  • 9马冠生,李艳平,潘慧,杜维婧,张倩,刘爱玲.177名北京城郊初中学生体脂含量的测定[J].营养学报,2003,25(4):353-356. 被引量:13
  • 10DEMPSTER P,A ITKENS S.A New Air Displacement Method for the Determination of Human Body Composition[J].Med Sci Sports Exe,1995,27(12):1692-1697.

二级参考文献24

  • 1Ziegler EE, Filer LJ. Present knowledge in nutrition [M]. Seventh edition, Washington DC: ILSI Press, 1996: 7-12.
  • 2Fields DA, Goran MI. [J]. J Appl Physiol,2000, 89: 613-620.
  • 3Dempster P, Aitkens S. [J]. Med Sci Sports Exerc, 1995, 27(12) : 1692-1697.
  • 4Nunez C, et al. [J]. Eur J Clin Nutr, 1999,53 : 382-387.
  • 5McCrory MA, et al. [J]. Med Sci Sports Exerc,1995, 27(12) : 1686-1691.
  • 6Iwaoka H, et al. [J]. J Nutr Sci Vitaminol,1998, 44: 561-568.
  • 7Miyatake N, et al. [J]. Diabetes, Obesity and Metabolism, 1999, 1: 347-351.
  • 8Well JCK, Fuller NJ. [J]. Int J Obes Relat Metab Disord, 2001, 25: 1161-1167.
  • 9Biaggi RR, et al. [J]. Am J Clin Nutr, 1999,69 : 898-903.
  • 10Collins MA, et al. [J]. Med Sci Sports Exert,1999, 31: 1350-1356.