用于高强度聚焦超声热消融的PAA模块:模块的制作及声学参数的测量(英文) 被引量:8

PAA phantom for use in thermal ablation of high intensity focused ultrasound: Phantom fabrication and acoustic parameters measurements
下载PDF
导出
摘要 目的研制一种透明的用于高强度聚焦超声热消融的PAA模块,并对其声学参数进行测量。方法观察模块浓度的不同所致颜色与透明度的变化,并对PAA模块的声学特性包括密度、声速及声衰减进行测量。结果(1)PAA模块为透明的琥珀色,蛋白浓度越高,颜色越深,透明度越低。(2)PAA模块的密度与水相近,为1.0250g/cm3至1.0617g/cm3。声速与模块浓度之间有明显的相关性。声衰减从0.125dB/cm至0.329dB/cm,随着模块密度与探头频率的增加而增加。结论PAA模块是一种较好的用于高强度聚焦超声热消融的均质仿体,实验结果有较好的重复性。 Objective An optically transparent phantom used for thermal ablation of HIFU has been developed with acoustic parameters measured. Methods The color and transparency of different concentration phantoms were observed. Properties including density, acoustic speed and acoustic attenuation of PAA gel phantom were measured. Results (1)PAA gel phantom was amber and transparent. The color became heavier and transparency decreased with the egg-white concentration increasing. (2)The density of PAA gel phantom was similar to water, which was from 1.0250g/cm3 to 1.0617g/cm3. There was a significant correlation between sound speed and phantom concentration. The acoustic attenuation was from O.125dB/cm to 0.329dB/cm, increased with the phantom density and transducer frequency increasing. Conclusions PAA gel phantom was homogeneous and a good mimicking phantom for HIFU thermal ablation. The experiment results had a satisfied reproducibility.
出处 《声学技术》 CSCD 北大核心 2006年第6期613-616,共4页 Technical Acoustics
关键词 聚丙烯酰胺(PAA) 模块 高强度聚焦超声(HIFU) 热消融 声学参数 polyacrylamide (PAA) phantom high intensity focused ultrasound (HIFU) therma ablation acoustic parameters
  • 相关文献

参考文献3

  • 1McDonald M,Lochhead S,Chopra R,et al.Multimodality tissue-mimicking phantom for thermal therapy[J].Phys Med Biol,2004,49(4):2767-2768.
  • 2Sakuma I,Takai Y,Kobayashi E,et al.Navigation of high intensity focused ultrasound applicator with an integrated three-dimensional ultrasound imaging system[J].LNCS,2002,2489(10):133-139.
  • 3Lafon C,Zderic V,Noble ML,et al.Gel phantom for use in high-intensity focused ultrasound dosimetery[J].Ultrasound Med Bio,2005,31(10):1383-1389.

同被引文献47

  • 1姜立新,胡兵.高强度聚焦超声作用下的组织温度场测量技术的研究进展[J].声学技术,2006,25(1):43-47. 被引量:5
  • 2Shariat SF,Raptidis G,Masatoschi M,et al.Pilot study of radiofrequency interstitial tumor ablation (RITA) for the treatment of radio-recurrent prostate cancer.Prostate,2005,65(3):260-267.
  • 3Wittkampf FH,Nakagawa H.RF catheter ablation:Lessons on lesions.Pacing Clin Electrophysiol,2006,29(11):1285-1297.
  • 4McDonald M,Lochhead S,Chopra R,et al.Multimodality tissue-mimicking phantom for thermal therapy.Phys Med Biol,2004,49(13):2767-2778.
  • 5Bouchard LS,Bronskii MJ.Magnetic resonance imaging of thermal coagulation effects in a phantom for calibrating thermal therapy devices.Med Phys,2000,27(5):1141-1145.
  • 6Lazebnik M,Madsen EL,Frank GR,et al.Tissue-mimicking phantom materials for narrowband and ultrawideband microwave applications.Phys Med Biol,2005,50(18):4245-4258.
  • 7Lobo SM,Afzal KS,Ahmed M,et al.Radiofrequency ablation:modeling the enhanced temperature response to adjuvant NaCl pretreatment.Radiology,2004,230(1):175-182.
  • 8Kuroda M,Kato H,Hanamoto K,et al.Development of a new hybrid gel phantom using carrageenan and gellan gum for visualizing three-dimensional temperature distribution during hyperthermia and radiofrequency ablation.Int J Oncol,2005,27(1):175-184.
  • 9Bu-Lin Z,Bing H,Sheng-Li K,et al.A polyacrylamide gel phantom for radiofrequency ablation.Int J Hyperthermia,2008,24(7):568-576.
  • 10Wiersma J,Van Dijk JD.RF hyperthermia array modeling; validation by means of measured EM-field diatributions.Int J Hyperthermia,2001,17(1):63-81.

引证文献8

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部