期刊文献+

A Point-of-Care Test System for Biochemical Blood Analysis Based on ADUC824

A Point-of-Care Test System for Biochemical Blood Analysis Based on ADUC824
下载PDF
导出
摘要 A point-of-care test system has been studied in this paper.It was used to determine substances in blood such as Hemoglobin (HB),Aspartate Aminotransferase (AST),Creatine Kinase (CK) and so on.Based on the principle of amperometric determination,the research on detecting weak current signals was carried on.At the same time as to the weak signals (nA level),magnifying,sampling and processing the signals were also studied.Controlled by ADUC824 and assisted by other units, every substance could be determined automatically and rapidly integrated with the corresponding biosensor.In the experiment, the minimum detectable current of the instrument (YT2005-1) is 0.2 nA.With regard to the 1 nA which the experiment demanded,it could be up to the mustard.And the system can provide results in 180 s with a long term stability. A point-of-care test system has been studied in this paper. It was used to determine substances in blood such as Hemoglobin(HB), Aspartate Aminotransferase (AST), Creatine Kinase (CK) and so on. Based on the principle of amperometric determination, the research on detecting weak current signals was carried on. At the same time as to the weak signals (nA level), magnifying, sampling and processing the signals were also studied. Controlled by ADUC824 and assisted by other units, every substance could be determined automatically and rapidly integrated with the corresponding biosensor. In the experiment, the minimum detectable current of the instrument (YT2005-1) is 0.2 nA. With regard to the 1 nA which the experiment demanded, it could be up to the mustard. And the system can provide results in 180 s with a long term stability.
出处 《稀有金属材料与工程》 SCIE EI CAS CSCD 北大核心 2006年第A03期75-79,共5页 Rare Metal Materials and Engineering
关键词 point-of-care test ADUC824 weak signals HB AST and CK point-of-care test ADUC824 weak signals HB, AST and CK
  • 相关文献

参考文献3

  • 1Lianshi Feng, Kaigang Li. The Measurement and Application of Common Physiological and Biochemical Parameter Using in Athlete Enginery Assess[M]. Beijing: The Publication of People's Physical Culture, China, 2002:46.
  • 2Liying Jiang, Xinxia Cai, Chunxiu Liu. JNanosci Nanotech[J], 2005, 5:1301.
  • 3Chunxiu Liu, Huaqing Li, Liying Jiang et al. Asian Conference on Nanoscience & Nanotechnology (Asian NANO 2004)[C]. Beijing: Asian NANO, 2004.