期刊文献+

逆向计划调强放射治疗的临床验证 被引量:1

Clinical Verification of Inverse Intensity Modulated Radiation Therapy
下载PDF
导出
摘要 目的:探讨逆向计划调强放射治疗IMRT临床治疗前的验证方法。方法:用胶片法验证Cadplan(包括Helios逆向计划系统)三维治疗计划系统生成的各个照射野注量图与在Varian23EX直线加速器(安装120叶MLC)照射中实际释放的注量图的一致性;用电离室法验证等中心绝对剂量计算值与实测值的符合性。结果:等中心绝对剂量的误差小于5%,符合临床要求;各射束轴垂直方向测得的注量图与计划系统计算的注量图一致;等中心位置验证显示,CT模拟的正侧位DRR片与加速器正侧位验证片误差在±2 mm以内。结论:Varian逆向调强放射治疗系统符合临床要求。 Objective: To evaluate Clinical verification of inverse intensity modulated radiation therapy(IMRT), Methods: To verify whether the fluence map of each beam desired by Helios inverse planning system was actually delivered in Varian 23 EX Linear accelerator with 120 leafes collimatorMLC, a piece of film under a homogeneous polystyrene phantom was irradiated vertically with each of the beams to record the deposited dose, which is the fluence map actually deliverd. The ionization chamber in a water phantom was used to measure the central point dose and another point dose. Results: The fluence maps measured by the film were well consistent with that predicted by the planning. The error between measured dose and predicted dose in the central poifit was less than 3%, but the error of dose in another point varied more widely. The film measurements of the isodose distributions (on axial plane) coincided with the predicted distributions. Verification on Accuracy of isocenter location was done by measurement and comparing the distance between the isocenter and anatomy mark in DRR films from CT simulator and verification films from Varian 23 EX Linear accelerator. The difference in setup is less than ±2 mm. Conclusions: the present investigation shows that the IMRT delivered by Varian 23 EX Linear accelerator with 120 leafs collimator(MLC) is available.
出处 《中国医学物理学杂志》 CSCD 2007年第1期4-6,共3页 Chinese Journal of Medical Physics
关键词 调强放射治疗 逆向计划 体模 临床治疗前的验证 intensity modulated radiation therapy inverse planning fluence map anthropomorphic phantom
  • 相关文献

参考文献4

  • 1Low DA, Gerber RL, Mutic S, et al. Phantoms for IMRT dose distribution measurement and treatment verification[J]. Int J Radiat Oncol Biol Phys, 1998,40:1231-1235.
  • 2Leybovich LB, Sethi A, Dogan N. Comparison of ionization chambers of various of volumes for IMRT absolute dose verification[J]. MedPhys, 2003,30(2):119-123.
  • 3李高峰,朱庙生,吴钦宏,高鸿,何京学,修霞.逆向计划调强适形放射治疗的质量保证[J].中华放射肿瘤学杂志,2002,11(3):190-193. 被引量:53
  • 4Mubata CD, Childs P, Bidmead AM.A quality assurance procedure, for the Varianmuhi-leaf collimator[J]. Phys.Med.Biol, 1997,42:423-431.

共引文献52

同被引文献6

  • 1陈勇,包尚联,黄斐增,袁克虹,肖桂平.HW-Plan放射治疗计划系统的实验验证[J].中国医学物理学杂志,2005,22(6):691-694. 被引量:8
  • 2马金利,蒋国梁,傅小龙,吴开良,李龙根.个体患者适形调强放疗计划模体内剂量实测验证技术的建立[J].中华放射医学与防护杂志,2007,27(1):80-84. 被引量:2
  • 3胡逸民.不均匀组织对剂量分布影响的校征方法//胡逸民,主编.肿瘤放射物理学.北京:原子能出版社.1999,204-205.
  • 4International Commission on Radiation Units and Measurements. ICRU Reports 42:use of computers in external beam radiotherapy procedures with high-energy photons and electrons. Bethesda:ICRU, 1987 : 1-70.
  • 5Radiation treatment planning dosimelry verification. Radiation Therapy Committee Task Group No. 23// American Association of physicists in Medicine. AAPM Report No. 55. New York: The American Institute of Physics Inc, 1995 : 1-11.
  • 6Petrie MP, Clark BG, Robar JL. A comparison of two commercial treatment planning systems to IMRT. J Appl Clin Med Phys,2005, 6:63-80.

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部