摘要 对ACL TOP全自动血凝分析仪进行全面性能评价,建立优化适用的血凝分析系统。评价结果表明ACL TOP血凝仪具有良好的重复性、稳定性、线性和抗生物性物质干扰能力,与ACL Futura血凝仪常规凝血指标比对结果不理想,需重新建立基于ACL TOP血凝分析系统的项目参考区间。 Evaluates the performance of ACL TOP automatic coagulation analyzer and establishes an optimized coagulation analytical system. The results show that ACL TOP coagulation analyzer has good repeatability, stability, linearity and high capability of anti - interference. However, the results of comparison to ACL FUTURA are not identical. So it is necessary to determine new reference range for tests on ACL TOP coagulation analytical system.
出处 《医疗设备信息》 2007年第2期48-51,共4页 Information of Medical Equipment
关键词 全自动分析系统 血液凝固 ACL TOP性能评价 automatic analytical system coagulation ACL TOP performance evaluation
  • 相关文献

参考文献5

  • 1NCCLS. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline,Second Edition [S ]. NCCLS document EP5-A2[ISBN 1-56238-542-9]. NCCLS, 940 WestValley Road, Suite 1400, Wayne, Pennsylvania 19087- 1898 USA, 2004.
  • 2杨昌国,许叶,张抗.线性评价和干扰试验中NCCLS评价方案的应用[J].临床检验杂志,1999,17(3):184-186. 被引量:165
  • 3NCCLS. Interference Testing in Clinical Chemistry; Approved Guideline[S]. NCCLS document EP7-A [ISBN 1-56238-480-5]. NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2002.
  • 4NCCLS. Method Comparison and Bias Estimation Using Patient Samples[S]; Approved Guideline- Second Edition. NCCLS document EP9-A2[ ISBN 1 - 56238- 472 -4]. NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2002.
  • 5张晓虹.CA510全自动血凝仪的使用与维护[J].医疗设备信息,2005,20(6):73-74. 被引量:11