摘要 采用微弧氧化法在钛合金(Ti-6Al-4V)基体上制备TiO2生物陶瓷涂层.试验研究了不同微弧氧化电压和脉冲频率对涂层微观形貌,涂层厚度及相组成的影响.结果表明:涂层厚度随氧化电压的升高而增加,随脉冲频率的增加而减小.SEM观察显示:涂层微孔数目随电压的升高而增加,随脉冲频率的增加表面粗糙度增大.XRD相结构分析表明:涂层中的主要相为金红石型TiO2和锐钛矿型TiO2及CaTiO3. TiO2 bioceramic coatings were prepared on the surface of Ti - 6Al - 4V by micro - arc oxidation (MAO) technique. The effects of different oxidation voltage and pulsed frequency on microstructure, thickness and phase constitute of the coatings have been investigated. The results demonstrate that with the oxidation voltage increasing, the thickness of the coatings increase and with the pulsed frequency increasing, the thickness of the coatings decrease. SEM indicates that with the oxidation voltage increasing, the number of micro-pore of the coatings increase and with the pulsed frequency increasing, the surface roughness of the coatings increase. XBD analysis indicates that the mainly phase are anatase TiO2, rutile TiO2 and CaTiO3.
出处 《佳木斯大学学报(自然科学版)》 CAS 2007年第1期52-55,共4页 Journal of Jiamusi University:Natural Science Edition
基金 黑龙江省自然科学基金项目资助(E0304) 黑龙江省研究生创新基金资助
关键词 微孤氧化 电参数 涂层 TIO2 micro-arc oxidation electrical parameters coating TiO2
  • 相关文献

参考文献8

二级参考文献16

  • 1陈治清.口腔种植界面的研究(Ⅳ)[J].生物医学工程学杂志,1989,6(4):312-318. 被引量:5
  • 2邓志威,薛文彬,汪新福,陈如意,来永春.铝合金表面微弧氧化技术[J].材料保护,1996,29(2):15-16. 被引量:91
  • 3[1]Mandl S, Sader R, Krause D, et al. Investigation on plasma immersion ion implantation treated medical implants [J]. Biomol Eng, 2002;(19): 129-132.
  • 4[2]Yerokhin AL, Voevodin AA, Lyubimov VV, et al. Plasma electrolytic fabrication of oxide ceramic surface layers for tribotechnical purposes on aluminium alloys [J]. Surf Coat Technol, 1998; (110): 140-146.
  • 5[3]Tian J, Luo ZZ, Qi SK, et al. Structure and antiwear behavior of micro-arc oxidized coatings on aluminum alloy [J]. Surf Coat Technol, 2002; (154): 1-7.
  • 6[4]Han Y, Hong SH, Xu KW. Synthesis of nanocrystalline titania films by micro-arc oxidation [J]. Mater Let, 2002; (56): 744-747.
  • 7[5]Hamada K, Kon M, Hanawa T, et al. Hydrothermal modification of titanium surface in calcium solutions [J]. Biomaterials, 2002; (23): 2265-2272.
  • 8[6]Milena F, Alberto C, Gianni R, et al. In vitro and in vivo behaviour of Ca- and P-enriched anodized titanium [J]. Biomaterials, 1999; (20): 1587-1594.
  • 9[7]Lange R, Lüthen F, Beck U, et al. Cell-extracellular matrix interaction and physico-chemical characteristics of titanium surfaces depend on the roughness of the material [J]. Biomol Eng, 2002; (19): 255-261.
  • 10[8]G(O..)ransson A, Jansson E, Tengvall P, et al. Bone formation after 4 weeks around blood-plasma-modified titanium implants with varying surface topographies: An in vivo study [J]. Biomaterials, 2003; (24): 197-205.

共引文献161

同被引文献68

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部