摘要
作者提出了一种应用联合国原子辐射效应科学委员会1982年报告中推荐的放射工作人员个人剂量监测结果的几个参量[1]:个人年平均剂量当量、年集体剂量当量和个人剂量当量大于15mSv人员年集体剂量当量与总年集体剂量当量的比值MR15的新方法。并以这种方法评价上海市放射工作人员外照射个人剂量水平。结果表明在1990~1994年期间上海市放射工作人员个人剂量[2]年平均剂量当量在0.40~4.04mSv,低于国家标准所规定的剂量当量限值的十分之一(5mSv),超过年剂量当量的十分之三(15mSv)有116人次,MR15在0.01~0.49。有328人次超调查水平(月剂量当量限值1.20mSv).对超调查水平者进行随访调查。针对"真实"、"虚假"者进行详细分析,并为有关放射工作单位提出相应的改进意见和个人防护措施。并指出今后搞好放射防护和个人剂量监测的关键[3]。
Anew method used for estimating several parameters of personal dose monitoring of radiation staffs,which were recommended by the UNSCEAR in 1982(the individual annual average dose equivalent,the annual collective dose equivalent and the ratio of the annual collective dose eqaivalent delivered at individual dose exceeding 15 mGy per year to the total annual collective dose),MR15 was proposed and has been employed to evaluate the level of the personal external exposure dose to radiation staff in Shanghai.The annual average dose equivalent for radiation workers was 0.40-0.404mSv,which was less than 1/10 of the national criterion(5mSv);however,the person-times exposed to more than 3/10 of the criterion were 116,MR15 being 0.01-0.49.328 person-times surpassed the investigation level(the limit value of the monthly dose is 1.20mSv).Those people were specially paid visits and“true”or“false”was identified in detail.Moreover,both personal protective measures and other aspects to be improved were suggested.The key points of the improvement in personal protection and of personal dose monitoring were commented.
出处
《中华放射医学与防护杂志》
CSCD
北大核心
1996年第4期264-267,共4页
Chinese Journal of Radiological Medicine and Protection