酸性氧化电位水与戊二醛消毒内镜临床观察 被引量:32

Disinfection of Endoscope by EOW and Glutaraldehyde:A Clinical Observation
下载PDF
导出
摘要 目的 了解酸性氧化电位水(EOW)和戊二醛用于临床无痛胃镜和肠镜消毒的效果、对人体危害及稳定性。方法 随机抽取门诊和住院患者做无痛胃镜和肠镜检查污染前后胃、肠镜,采用戊二醛消毒液为对照,比较两者的消毒效果、对人体的危害性及稳定性进行临床观察。结果 EOW消毒胃、肠镜对常见病原菌1min杀灭率为93.75%~98.22%、3min为99.30%~99.95%、5min为100.00%;戊二醛消毒胃、肠镜3min杀灭率为81.20%~89.29%、5min为90.38%~93.04%、10min为99.92%~99.96%;EOW对人体无过敏反应和黏膜刺激,但稳定性差;而戊二醛对人体均有黏膜刺激和个别过敏反应。结论 EOW3~5min可杀灭胃、肠镜检查中的病原菌,戊二醛消毒〉10min才能杀灭常见病原菌;EOW杀菌作用强,快捷、高效、安全,利于环保,但必须现生成现用;使用戊二醛必须注意个人防护。 OBJECTIVE To know the disinfection efficacy of electrolyzed oxidizing water (EOW) and glutaraldehyde on clinical used gastroscope and enteroscope in our hospital and to analyze their stability and probable harm. METHODS The gastroscope and enteroscope before and after use on randomly selected outpatients and inpatients were examined on their contamination, and the efficacy of EOW was compared with that of glutaraldehyde. RESULTS After using EOW and glutaraldehyde to disinfect the gastroscope and enteroscope the former showed that its killing rate on commonly encountered pathogens for 1 min was 93.75-98.22%, for 3 min was 81.20- 89. 29%, and for 5 min was 100.00% ; the latter showed that its killing rate for 3 min was 81.20-89. 29%, for 5 min was 90.38-93. 04%, and for 10 min was 99.92-99. 96%. EOW was non-irritative to mucosa, and didnrt cause allergic reaction, but its stability was poor. The glutaraldehyle could bring some side effects, such as some allergy, and mucosa stimulation. CONCLUSIONS It only needs 3-5 min for EOW to kill polluted bacteria on gastroscope and enteroscope to attain the disinfection eligibility, but glutaraldehyde needs more than 10 rain to get to the disinfection eligibility. The effect of EOW is strong, rapid, active and environment-safe, it is worthy of application, but it must be applied immediately as soon as possible after production. The use of glutaraldehyde must strengthen the measures of personal protection.
出处 《中华医院感染学杂志》 CAS CSCD 北大核心 2007年第3期292-294,共3页 Chinese Journal of Nosocomiology
关键词 酸性氧化电位水 戊二醛 消毒 内镜 临床 观察 EOW Glutaraldehyde Disinfection Endoscope Harm Observation
  • 相关文献

参考文献5

二级参考文献34

  • 1冼日凤,周国萍,余玲,林祖坚,黄彦.连台手术内镜快速灭菌方法的应用研究[J].中华医院感染学杂志,2004,14(1):63-63. 被引量:21
  • 2张艳红,朱莉莉,许迎红,俞幼知.氧化电位水在胃镜消毒中的应用研究[J].中华医院感染学杂志,2004,14(4):415-416. 被引量:19
  • 3王健刚,刘正霞,朱杰平,陶萍.擦立净消毒胃镜效果观察[J].中国消毒学杂志,1995,12(1):41-42. 被引量:5
  • 4刘胜文.实用医院感染管理手册[M].北京:北京医科大学中国协和医科大学联合出版社,1991.195-197.
  • 5Alfa MJ,Sitter DL. In hospital evaluation of ortho-phthal aldehyde as a high level disinfectant for flexible endoscopes[J].J Hosp Infect, 1994,26:15.
  • 6Axon A, Kruse A, Urgell R,et al. European Society of Gastrointestinal Endoscopy; guidelines on cleaning and disinfection in gastrointestinal endoscopy[J]. Endoscopy, 1995, 27: 199-202.
  • 7The Hong Kong Society of Endoscopy Nurses. Recommended standard for reprocessing of flexible endoscopy and accessories [S]. 2000. 1-7.
  • 8Alvarado CJ, Reichelderfer M, APIC. APIC Guideline for infection prevention and control in flexible endoscope[J]. Am J Infect Control, 2000, 28:138-55.
  • 9.医院消毒技术规范[S].[S].中华人民共和国卫生部,2002.171-172,266.
  • 10Babb J R,Bradley C R.The mechanics of endoscope disinfention[J].J Hosp Infect,1991,18(Supplement A):130.

共引文献285

同被引文献280

引证文献32

二级引证文献176

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部