血管支架筋宽误差分析及激光加工工艺优化 被引量:4

Width Error Analysis of Stent Strut and Optimization of Laser Cutting Procedure
下载PDF
导出
摘要 为解决血管支架激光切割中普遍存在筋宽不均匀的问题,通过激光切割工艺试验及对管材的运动分析,得出影响筋宽的主要因素包括系统误差造成的管材切割点切向速度变化和激光离焦量.根据研究结果,优化了激光切割参数.限制管材回转偏心在0.030 mm以下,并设置激光离焦量为-0.05 mm(设备显示值),筋宽变化幅度控制在0.010 mm以内,满足设计要求. To solve the problem of uneven strut widths of stents cut by laser, laser cutting procedure test was conducted, and the motion of the tube was analyzed. The results show that the deviation of the strut width is mainly caused by the variation in laser defocus and that in the tangent velocity at the cutting point. The procedure parameters were optimized according to the above analysis. The strut widths of the stent cut with the optimized procedure vary within 0. 010 mm, conforming to the design specifications, when the eccentricity of tube at cutting point is within 0. 030 mm and the defocus is -0.05 mm.
出处 《西南交通大学学报》 EI CSCD 北大核心 2007年第2期229-233,共5页 Journal of Southwest Jiaotong University
基金 国家重点基础研究规划项目(973)(2005CB623904) 国家自然科学基金资助项目(30318006)
关键词 激光切割 筋宽 工艺参数 工艺试验 优化 血管支架 laser cutting strut width procedure parameter procedure test optimization stent
  • 相关文献

参考文献6

  • 1王涛,古桂茹,李玉翔,姚建铨,刘宝利,蔡彬晶.激光在机械制造技术中的应用[J].激光杂志,2006,27(3):75-76. 被引量:12
  • 2ETAVE F,FINET G,MAURICE B,et al.Mechanical properties of cornary stent deternined by using finite element analysis[J].Journal of Biomechanics,2001,34:1 065-1 070.
  • 3杨杰,黄楠,杜全兴,张勇,陈虬.描述血管支架力学性能的新概念──形状系数[J].生物医学工程学杂志,2006,23(1):74-77. 被引量:10
  • 4ROGERS C,TSENG D Y,SQUIRE J C,et al.Balloon-artery interactions during stent placement:a finite element analysis approach to pressure,compliance,and stent design as contributors to vascular injury[J].Circ.Res.,1999,84(4):378-385.
  • 5张康年.激光加工技术[M].北京:化学工业出版社,2004:38.
  • 6MCGARRY J P,O DONNELL B P,MCHUGH P E,et al.Analysis of the mechanical performance of a cardiovascular stent design based on micromechanical modeling[J].Computation Materials Science,2004,31:421-438.

二级参考文献10

  • 1兰信钜 等.激光技术[M].湖南:湖南科学技术出版社,1979.7.
  • 2Francesco Migliavaccaa, Lorenza Petrini, Valeria Montanari, et al. A predictive study of the mechanical behaviour of coronary stents by computer modeling. Medical Engineering & Physics,2005; 27:13.
  • 3David Chua SN, Mac Donald BJ, Hashmi MSJ. Finite element simulation of stent and balloon interaction. Journal of Materials Processing Technology, 2003;143: 591.
  • 4James E Moore, Joel L Berry. Fluid and solid mechanical implications of vascular stenting. Annals of Biomedical Engineering, 2002;30: 498.
  • 5Auricchio F, Di Loreto M, Sacco E. Finite-element analysis of a stenotic artery revascularization through a stent insertion. Comp Meth Biomech Biomed Eng, 2001; 4 : 249.
  • 6Dumoulin C, Cochelin B. Mechanical behaviour modelling of balloon-expandable stents, Journal of Biomechanics, 2000; 33: 1461.
  • 7王仁.塑性力学[M].北京:科学出版社,1998.46-57.
  • 8Francesco Migliavaccaa, Lorenza Petrinib, Maurizio Colombo,et al. Mechanical behavior of coronary stents investigated through the finite element method, Journal of Biomechanics,2002; 35 : 803.
  • 9Etave F, Finet G, Maurice Boivin, et al. Mechanical properties of coronary stents determined by using finite element analysis.Journal of Biomechanics,2001 ; 34:1065.
  • 10王涛 李喜福 郭玲.激光动态识别Line元件三维形状及动态研究[J].中国激光,2002,29(6):608-610.

共引文献20

同被引文献85

引证文献4

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部