气相色谱-常压微波等离子体发射光谱系统操作参数的考察及最优化

Investigation and Optimization of the Operating Parameters of Gas Chromatography-Atmospheric Pressure Helium Microwave-Induced Plasma Emission Spectrometry
下载PDF
导出
摘要 本文对影响GC-MIP检测性能的操作参数(包括等离子体观测位置,微波功率和He支持气流速)进行了最优化研究。结果表明,对C、H、D、O、S、P、F、Cl、Br和I元素来说,每个元素都有各自的一套最佳操作参数。此外,还对记忆效应进行了讨论。 The operating parameters, including observing position, microwave power and helium flow rate, which have significant effects on the detecting performance of gas chromatography atomospheric pressure helium microwave-induced plasma emission spectrometry system, have been optimized for the determination of carbon, hydrogen, deuterium, oxygen, sulfur, phosphorus, fluorine, chlorine, bromine and iodine. The memory and wall effect are discussed as well.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1990年第9期812-815,共4页 Chinese Journal of Analytical Chemistry
关键词 GC-MIP 操作参数 最优化 Microwave-induced plasma atomic emission spectrometry Optimization of operating parameters Capillary gas chromatography
  • 相关文献

参考文献3

  • 1贾崇荣,分析仪器,1989年,4卷,34页
  • 2曾克慰,色谱,1987年,5卷,1期,35页
  • 3曾克慰,化物报道,1979年,1卷,119页