摘要
目的:了解多频听觉稳态诱发反应(MASSR)对骨导客观听阈及对传导性聋评估的准确性。方法:对30例传导性聋患者进行气、骨导MASSR及纯音测听,并将MASSR反应阈和纯音听阈进行比较。结果:气、骨导MASSR反应阈和纯音听阈之间具有较好的相关性,MASSR对传导性聋评估的准确性和纯音测听相似。结论:MASSR可用作传导性聋患者客观听阈的评估。
Objective : To assess the accuracy of threshold using the multiple auditory steady state responses (MASSR) to bone conduction stimuli in a group of patients with conductive hearing loss. Method: Thirty patients with conductive hearing loss were assessed by MASSR and by pure-tone audiometry, and compared the results of these two methods. Result:The thresholds of MASSR and those of PTA were significantly correlated, and the accuracy of MASSR were similar to PTA for the patients with conductive hearing loss. Conclusion: MASSR can be used to assess the hearing thresholds of patients with conductive hearing loss.
出处
《临床耳鼻咽喉头颈外科杂志》
CAS
CSCD
北大核心
2007年第7期293-295,共3页
Journal of Clinical Otorhinolaryngology Head And Neck Surgery
关键词
多频听觉稳态诱发反应
测听法
纯音
听力损失
Multiple auditory steady-state responses
Pure-tone audiometry
Hearing loss,conductive