停流-催化分光光度法测定痕量锰 被引量:2

Stopped-Flow Catalytic Spectrophotometric Determination of Trace Manganese
下载PDF
导出
摘要 本文用流动注射分析仪,采用停留技术,建立了基于以氨三乙酸(NTA)为助催剂、Mn^(2+)对KIO_4氧化孔雀绿的催化作用测定痕量锰的催化光度法,在0~7.2ng/ml Mn^(2+)呈线性,检测限为0.07ng/ml。用于江湖水中痕量锰的测定,获得满意结果。 A stopped-flow catalytic spectrophotometric method for determining Mn by using a flow injection analyzer has been established. The method is based on the catalytic effect of Mn^(2+)on malachite green-KIO_4 in the presence of nitrilo triacetic acid as an acceterator. A linear range is obtained from 0 to 7. 2ng Mn/ml, and the detection limit of Mn is 0. 07n g/ml. The proposed method has goad selectivity, and has been applied to the determination of Mn river water with satisfactoly results. The relative standard deviation is smaller than 3. 9%.
作者 李开雄 任英
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1990年第11期1041-1043,共3页 Chinese Journal of Analytical Chemistry
基金 自然科学基金
关键词 孔雀绿 KIO4 催化光度法 停流 Manganese Malachite green Potassium periodate Catalytic spectrophotometry Stopped-flow
  • 相关文献

参考文献3

  • 1王尊本,分析化学,1987年,15卷,2期,145页
  • 2赵宏樵,环境化学,1987年,6卷,3期,73页