摘要
本文提出在0.05mol/LKCI-0.02mol/LHNO3-0.0025mol/L(CH2)5N4-3×10(-5)mol/LHgCl2体系中,采用微分脉冲阳极溶出伏安技术测定大气中痕量镉、铅,该方法灵敏度高,检测下限达5.0×10(-11)ug/ml(即0.05ppb),样品回收率均在95-105%范围内,十次平行测定结果的相对标准偏差不大于5%。
The article put a new method for determining the Cadmium and lead in air byAnodic Stripping Differential Pulse voltiammetry, in system, 0. 05mol/L KCI-0. 02mol/LHNO3 - 2. 5 ×10-3mol/L (CH2)5N4 -3 ×10-5mol/L HgCl2.This method is sensitive and accu-rate. It's detection limit reach the 5 ×10-11ug/ml (0.05ppb). The recovery rate of samples iscontrolled in 95-105%. It's less than 5% that relative standard deviations of measurmet re-sults of ten time.
关键词
微分脉冲
大气
镉
铅
测定
阳极溶出伏安法
Anodic Styipping Differential Pulse Voltammetry, Air, Lead, Cadmium