不同灌注软件对正常人肾脏灌注的比较研究 被引量:4

Study of CT perfusion in normal kidneys
下载PDF
导出
摘要 目的:评价正常人肾脏CT灌注成像的可行性和准确性,分析不同灌注软件测得数据的异同。方法:采用Siemens 64层螺旋CT机对24例没有肾脏疾病的受检者进行双肾灌注扫描。原始数据分别经Siemens、Philips及GE灌注软件进行处理,对双侧肾脏灌注指标进行比较,同时对不同灌注软件测得的同一灌注参数进行比较。结果:左肾与右肾皮质的同一灌注指标无明显差异;不同软件测得的主动脉峰值、达峰时间,肾皮质的最大增强值和达峰时间之间无显著性差异,而肾皮质血流量之间和血容量之间有显著差异。结论:肾脏CT灌注成像能够准确地反映出肾脏的血流灌注特点,不同软件测得的数据在一定程度上具有一致性。 Objective: To assess the feasibility and accuracy of CT perfusion of normal human kidneys and compare the perfusion parameters acquired with different perfusion softwares. Methods: Twenty-four healthy subjects without renal problems underwent CT perfusion on a 64-slice CT of Siemens. The raw data was calculated with software of Siemens, Philips and GE respectively and the perfusion characteristic parameters acquired were compared. The perfusion parameters of bilateral kidneys were also compared. Results: There was no significant difference in the same parameter between bilateral kidneys. There was also no difference in aortic peak aortic enhancement(PAE) and time to peak(TTP), renal peak enhancement(PE) and TIP among different softwares. Obvious difference existed in renal blood flow (BF) and blood volume (BV) among different softwares. Conclusion: CT perfusion is a feasible and accurate technique to assess renal blood perfusion. Perfusion parameters acquired with different softwares have concordance to some extent.
出处 《中国临床医学影像杂志》 CAS 北大核心 2007年第3期174-177,共4页 Journal of China Clinic Medical Imaging
关键词 软件 kidney software
  • 相关文献

参考文献9

二级参考文献60

  • 1孙建男,么刚,李立军,郭启勇.肾脏CT灌注成像[J].中国医学影像技术,2004,20(8):1300-1302. 被引量:13
  • 2周剑,高培毅.动态CT灌注成像临床应用中的质量控制[J].中国医学影像技术,2003,19(11):1576-1578. 被引量:6
  • 3范光明,郭启勇,解学乾,陈丽英,于涛,侯阳.造影剂量和注射流率对肝脏CT灌注成像影响的实验研究[J].中国医学影像技术,2005,21(1):14-17. 被引量:14
  • 4[1]Eastwood JD,Lev MH,Azhari T,et al. CT perfusion scanning with deconvolution analysis:pilot study in patients with acute middle cerebral artery stroke [J]. Radiology,2002,222(1):227-236.
  • 5[2]Furukawa M,Kashiwagi S,Matsunaga N,et al. Evaluation of cerebral perfusion parameters measured by perfusion CT in chronic cerebral ischemia: comparison with Xenon CT [J]. J Comput Assist Tomogr,2002,26(2):272-278.
  • 6[3]Miles KA,Heyball MP,Dixon AK,et al. Functional images of hepatic perfusion obtained with dynamic CT [J]. Radiology,1993,188(2):405-411.
  • 7[4]Blomley MJK,Coulden R,Dawson P,et al. Liver perfusion studied with ultrafast CT [J]. J Comput Assist Tomogr,1995,19(3):424-433.
  • 8[5]Materne R,Van Beers BE,Smith AM,et al. Non-invasive quantification of liver perfusion with dynamic computed tomography and a dual-input one-compartmental model [J]. Clin Sci,2000,99(6):517-525.
  • 9[6]Cuenod CA,Leconte I,Siauve N,et al. Early changes in liver perfusion caused by occult metastases in rats: detection with quantitative CT [J]. Radiology,2001,218(2):556-561.
  • 10[7]Miles KA. Measurement of tissue perfusion by dynamic computed tomography [J].Br J Radiol,1991,64(761):409-412.

共引文献126

同被引文献73

引证文献4

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部