动态轮廓眼压计与Goldmann压平眼压计的临床应用比较 被引量:2

Clinical Comparison of the Dynamic Contour Tonometry to the Goldmann Applanation Tonometry
原文传递
导出
摘要 目的了解动态轮廓眼压计的临床实用性。方法67例患者134眼随即先后用动态轮廓眼压计与Goldmann压平眼压计进行眼压测量,比较两种眼压计测量值的差异。结果134眼中,用Goldmann压平眼压计测量的眼压平均值为(15.82±4.88)mmHg,用动态轮廓眼压计测量的眼压平均值为(14.5±4.33)mmHg,;两者无明显差异。但73%的患者的动态轮廓眼压计的眼压值高于Goldmann压平眼压计的测量值。两者的相关性为0.83。结论动态轮廓眼压计可以较好的反映患者的眼压,操作简便,值得推广应用。 Objeetive To compare the Dynamic Contour Tonomctry (DCT) with the Goldmann Applanation Tonomctry(GAT) in healthy subject.s. Methods 67 case (134 eyes) were random measured Intraocular Pressure (IOP) with the DCT and the GAT. Result In 134 eyes, The average IOP measured by GAT were 14.5+/-4.33mmHg, measured by DCT were 15.82+/-4.88mmHg, But 73% IOP of DCT were higher than GAT. The correlation coefficient between the IOP of two instruments was 0.83. Conelusion DCT can accuracy measure IOP of patients, easy to operate. It is valuable to be applied to clinical.
出处 《中国实用眼科杂志》 CSCD 北大核心 2007年第2期163-164,共2页 Chinese Journal of Practical Ophthalmology
关键词 眼匪 动态轮廓眼压计 GOLDMANN压平眼压计 Intraocular pressure Dynamic Contour Tonometry, Goldmann Applanation Tonometry
  • 相关文献

参考文献3

  • 1Kanngiesser HE Robert YCE Dynamic Contour Tonometry (ARVO abstract) Invest Opthalmol Vis Sci 2002;43,E°≤abstract°≥
  • 2Claude Kaufmann,Lucas M.Bachmann,Michael A Tbiel Invest Opthalmol Vis Sci 2003.VOL.44 3790-4
  • 3Christoph Kniestedt,Michelle Nee,Robert l,Arch Ophthalmol 2004;1122:1287-93

同被引文献34

  • 1乔荣华,刘丽娜,李美玉,任泽钦.Tono-Pen眼压计临床应用评价[J].中国实用眼科杂志,2004,22(12):980-982. 被引量:12
  • 2王铮,邱平,杨斌,钟兴武,张哓哓,刘泉,黄国富,周胜.动态眼压测量计与压平式眼压计测量LASIK手术前后眼压分析[J].中国实用眼科杂志,2006,24(5):472-474. 被引量:5
  • 3陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:217
  • 4李风鸣.眼科全书[M].北京:人民卫生出版社,1996.124-168.
  • 5Whitaere MM, Stein R. Sources of error with use of Goldmann- type tonometers. Surv Ophthalmol, 1993,38 : 1-30.
  • 6Kanngiesser HE, Kniestedt C, Robert YC. Dynamic contour to- nometry: presentation of a new tonometer. J Glaucoma,2005,14: 344 -350.
  • 7Duba I, Wirthlin AC. Dynamic contour tonometry for post-LASIK intraocular pressure measurements. Klin Monatsbl Augenheilkd, 2004,221:347-350.
  • 8Shimmyo M, Ross A J, Moy A, et al. Intmocular pressure, Goldma- nn applanation tension, corneal thickness, and comcal curvature in Caucasians, Asians, Hispanics, and African-Americans. Am J Ophthalmol,2003 ,136 :603- 613.
  • 9Doyle A, Lachkar Y. Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness. J Glaucoma ,2005,14:288-292.
  • 10Kamppeter BA, Jonas JB. Dynamic contour tonometry for intraocu- lar pressure measurement. Am J Ophthalmol,2005,140:318-320.

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部