期刊文献+

新型点接触锁定加压接骨板的研制——两种有限接触接骨板生物力学对照研究 被引量:2

A new point contact-locking compression plate design——The biomechanical evaluation of two limiting contact bone plate
下载PDF
导出
摘要 目的研制一种新型点接触锁定加压接骨板(PC-LCP),对其固定新鲜山羊胫骨表面形态特征及生物力学特性进行评价。方法应用压力敏感膜,比较新型点接触锁定加压钢板(PC-LCP)与有限接触加压接骨板(LC-DCP)固定新鲜山羊胫骨时接触面积、压力及应力变化。同时使用山羊新鲜配对胫骨15对,做成横断的骨折模型,分别用PC-LCP及LC-DCP加压固定,在多功能力学试验机上进行四点弯曲实验、扭转实验及垂直压缩试验。结果PC-LCP与LC-DCP固定新鲜山羊胫骨时接触面积、压力及应力差异有统计学意义(P<0.01)。PC-LCP固定骨折抗扭刚度略小于LC-DCP,抗弯刚度、垂直压缩刚度均略大于LC-DCP,但三者之间差异无统计学意义(P>0.05)。结论PC-LCP明显减少了与骨面的接触面积,同时固定骨折的生物力学性能达到临床需求。 Objective To determine the interface contact profiles and the biomechanical of the new design plate PC-LCP in fixing the fresh sheep tibia. Methods Using the Pressure sensitive film and the image analysis system to evaluated the variance of the interface characteristics of pressure,force and contact area when fixing the fresh sheep tibia. In each 15 pairs fresh sheep cadaveric tibia fixed by PC-LCP or LC-DCP, the Four point bonding,torsion tes tand axial compression was conducted to compare the bonding and torsion stiffness.Results The new design bone plate PC-LCP provide a significant differences in interface conact when compared to the LC-DCP (P 〈 0.01 )and there were no significant differences in bending, torsion stiffness and axial compression between the two experimental bone plates. Conclusion PC-LCP provided a advantage in deducing the contact area than LC-DCP and it attained the clinical demand of the biomechanical stability.
出处 《四川医学》 CAS 2007年第4期355-357,共3页 Sichuan Medical Journal
基金 重庆市科技攻关重大项目(CSTC 2006AA5014-1)
关键词 接骨板 锁定加压 点状接触 生物力学 bone plate locking compress point contact mechanical properties
  • 相关文献

参考文献9

  • 1Gerber C,Mast JW,Ganz R.Biological internal fixation of fractures[J].Arch Orthop Trauma Surg,1990,109(6):295~303
  • 2赵玉峰,李起鸿,顾祖超,许建中,杨柳.新型点状接触动力加压接骨板的研制(生物力学部分)[J].医用生物力学,2002,17(1):58-62. 被引量:11
  • 3Perren S.Evolution of the internal fixation of long bone fractures.The scientific basis ofbiological internal fixation:choosing a new balance between stability and blology[J].Bone Joint Surg Br,2002,84(8):1093~1110
  • 4Cordey J,Borgeaud M,Perren S.Force transfer between the plate and the bone:relative importance of the bending stiffness of the screws friction between plate and bone[J]. Injury,2000,31(Suppl3):21~28
  • 5Miclau T, Martin RE.The evolution of modern plate osteosynthesis[J].Injury,1997,28(suppl):3~6
  • 6吴雪晖,李起鸿,杨柳,陈锡秋.锥状点式接触钢板内固定对局部皮质骨微循环影响的实验研究[J].中华骨科杂志,2000,20(3):137-141. 被引量:50
  • 7赵玉峰,李起鸿,顾祖超,王爱民.新型点式接触动力加压接骨板固定对羊完整胫骨皮质骨血流量的影响[J].中华创伤杂志,2003,19(8):470-473. 被引量:19
  • 8John R,Fielda J,Rohan Edmonds-Wilsonb,et al.An evaluation of interface contact profiles in two low contact bone plates[J].Injured,2004,35(6):551~556
  • 9Lobo J,Shekhman M,Mishra S,et al.Effect of cortical contact area on biomechanical properties of LC-DCP plates for fixation of transverse fractures[J].Proceedings of AAOS,2001,68(6):339~341

二级参考文献20

  • 1王友,王震宇.接骨板内固定对局部皮质骨微循环影响的形态学研究[J].中华外科杂志,1993,31(4):248-250. 被引量:16
  • 2戴戎,戴闵,王开友,薛文东.应力松弛接骨板对骨几何形态和力学性能影响的实验观察[J].中华医学杂志,1995,75(7):414-416. 被引量:18
  • 3朱振安,戴克戎,裘世静,徐秀林,薛文东.坚硬接骨板固定及取出对骨力学性能的影响[J].中华创伤杂志,1997,13(1):21-23. 被引量:12
  • 4Jacobs RR, Rahn BA, Perren SM. Effects of plates on cortical bone perfusion. J Trauma, 1981, 21 : 91 -95.
  • 5Gunst MA. Interference with bone blood supply through plating of intact bone. In: Uhthoff HK, ed. Current concepts of internal fixation of fractures. 1nd ed. Berlin: Springer, 1980. 268 -276.
  • 6Gautier E, Rahn BA, Perren SM. Vascular remodeling. Injury,1995, 26 Suppl 2:11 - 19.
  • 7Perren SM, Cordey J, Rahn BA, et al. Early temporary porosis of bone induced by internal fixation implants. A reaction to necrosis, not to stress protection? Clin Orthop, 1988, (232) : 139 -151.
  • 8Danis R. The aim of internal fixation, Clin Orthop, 1979, (138) :23 - 25.
  • 9Uhthoff HK, Foux A, Yeadon A, et al. Two processes of bone remodeling in plated intact femora: an experimental study in dogs. J Orthop Res, 1993, 11:78-91.
  • 10Perren SM. The concept of biological plating using the limited contact-dynamic compression plate ( LC - DCP). Scientific background,design and application. Injury, 1991,22 Suppl 1 : 1 -41.