期刊文献+

基于传统轴流泵设计理论的人工心脏泵及其数值模拟 被引量:3

Applications of traditional axial flow pump design theory on artificial heart and CFD simulation
下载PDF
导出
摘要 传统轴流泵设计理论主要用来设计大型工业泵,很多都是经验设计的总结,人工心脏泵的体积相比较而言则要小得多。本文改进了传统轴流泵的设计理论,结合人工心脏泵的工作条件和水力条件,引入了扬程范围判断和速度三角形修正相结合的方法,配合迭代逼近获得了人工心脏泵的初始模型,CFD模拟分析为模型参数选择提供了依据。结论表明,这种设计方法可以用来快速简单的获得一系列高效的人工心脏泵模型,具有一定的代表性。 Traditional axial flow pump design theory accumulated with design experience was mainly used for industrial pump while artificial pump is much smaller than industrial pump. Initial heart pump model was obtained by improving on traditional theory combined with the working situation and hydraulic situation for artificial heart. Pump head range and velocity triangle was used with iteration approach for initial model and CFD simulation was applied for parameters decision. The conclusion shows that total design procedure can easily design series of high efficiency artificial heart pumps.
出处 《北京生物医学工程》 2007年第2期113-116,共4页 Beijing Biomedical Engineering
关键词 传统轴流泵设计理论 人工心脏泵 初始模型 CFD模拟分析 traditional axial pump theory artificial heart pump initial model CFD simulation
  • 相关文献

参考文献8

  • 1Yukiyasu Sezai.Progress and Future Perspectives in Mechanical Circulatory Support.Artificial Organs,2001,25 (5):318-322
  • 2[波兰]A.T.特罗斯科兰斯基.叶片泵计算与结构[M].机械工业出版社,1981.11.
  • 3Stepanoff AJ,ed.Centrifugal and Axial Flow Pumps:Theory,Design and Application,Second edition.Florida:Krieger Publishing Company,1992
  • 4Christopher Earls,Brennen.Hydrodynamics of Pumps.VT:Concepts NREC & Oxford University Press,1994
  • 5[苏]格染科,帕皮拉主编.叶片泵.吴仁荣,等译.北京:国防工业出版社,1982
  • 6Balje OE.Turbomachines:A Guide to Design,Selection,and Theory.New York:John Wiley & Sons Inc,1981
  • 7Apel J,Neudel F,Reul H.Computational Fluid Dynamics and Experimental Validation of a Microaxial Blood Pump.ASAIO J,2001,47:552-558
  • 8Apel J,Paul R,Klaus S,et al.Assessment of Hemolysis Related Quantities in a Microaxial Blood Pump by Computational Fluid Dynamics.Artificial Organs,2001,25:341-347

共引文献1

同被引文献26

  • 1云忠,谭建平,杨剑平.植入式微型螺旋血泵叶轮设计[J].机械工程与自动化,2006(5):4-5. 被引量:6
  • 2Yoshifumi Saijo, Hiroaki Okawai, et al. Evaluation of the Inner-Surface Morphology of an Artificial Heart by Acoustic Microscopy [J]. Artificial Organs, 2000, 24( 1 ) :64 -69
  • 3Allaire PE, Wood HG, et al. Blood Flow in a Continuous Flow Ventricular Assist Device [ J ]. Artificial Organs, 1999,23 ( 8 ) : 769 - 773
  • 4Chan WK, Wong YW, et al. Numerical Investigation of the Effects of the Clearance Gap Between the Inducer and Impeller of an Axial Blood Pump[ J]. Artificial Organs, 2005, 29 (3) :250 - 258
  • 5Arash Arvand, Nicole Hahn, et al. Comparison of Hydraulic and Hemolytic Properties of Different Impeller Designs of an Implantable Rotary. Blood Pump by Computational Fluid Dynamics [J]. Artificial Organs, 2004, 28(10): 892-898
  • 6[波兰]特罗斯科兰斯基,等著.耿惠彬译.叶片泵计算与结构.北京:机械工业出版社,1981
  • 7[苏联]格良科,帕皮拉,编.吴仁荣,等译.叶片泵.北京:国防工业出版社,1982
  • 8Thompson, Joe F,Ed. Numerical Grid Generation, North-Holland, 1982
  • 9AmyL Throckmorton, Alexandrina Untaroiu, et al. Computational Analysis of an Axial Flow Pediatric Ventricular Assist Device [ J]. Artificial Organs, 2004,28 (10) :881 - 891
  • 10Xinwei Song, Alexandrina Untaroiu, et al. Design and Transient Computational Fluid Dynamics Study of a Continuous Axial Flow Ventricular Assist Device[J]. ASAIO Journal,2004, 50:215 - 224