期刊文献+

Erythrocyte Fragment Count Predicts Hemolysis in Roller Pumps

Erythrocyte Fragment Count Predicts Hemolysis in Roller Pumps
下载PDF
导出
摘要 Objective: Hemolysis in blood pumps has been measured by various in vitro test methods, in which normalized index of hemolysis (NIH) was established. As NIH is complicated and difficult to calculate, erythrocyte fragment count is proposed in the present study to predict hemolysis in roller pumps. Methods: Five paired in vitro tests were conducted using the POLYSTAN pediatric pump(group A) and COBE pump( group B). Ten whole blood samples (400 ml ) were circled in the roller pump for 16 h. Erythrocyte fragments count and plasma-free hemoglobin (FHb) were measured before pumping and every two hours through circulation after four-hour-pumping. The morphological changes of erythrocyte were observed by scanning electron microscope. Results: The two groups' EFC and FHb levels were increased linearly during a long duration of pumping and linear regression of erythrocyte fragments count and plasma-free hemoglobin were correlated. Conclusion: Erythrocyte fragments count could be used as an index in evaluating the in vitro hemolytic properties of blood pumps. Objective: Hemolysis in blood pumps has been measured by various in vitro test methods, in which normalized index of hemolysis (NIH) was established. As NIH is complicated and difficult to calculate, erythrocyte fragment count is proposed in the present study to predict hemolysis in roller pumps.Methods: Five paired in vitro tests were conducted using the POLYSTAN pediatric pump(group A) and COBE pump(group B).Ten whole blood samples (400 m1) were circled in the roller pump for 16 h. Erythrocyte fragments count and plasma-free hemoglobin (FHb) were measured before pumping and every two hours through circulation after four-hour-pumping. The morphological changes of erythrocyte were observed by scanning electron microscope.Results: The two groups’EFC and FHb levels were increased linearly during a long duration of pumping and linear regression of erythrocyte fragments count and plasma-free hemoglobin were correlated.Conclusion: Erythrocyte fragments count could be used as an index in evaluating the in vitro hemolytic properties of blood pumps.
出处 《Chinese Journal of Biomedical Engineering(English Edition)》 2007年第1期33-38,共6页 中国生物医学工程学报(英文版)
关键词 Erythrocyte fragments count Roller pump HEMOLYSIS 滚压泵 溶血 预测 红细胞碎片计数法 人工血泵
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部