两种电子根尖定位仪测量准确性的实验研究 被引量:14

Evaluation of the Accuracy of Two Electronic Apex Locators: an in vitro Study
下载PDF
导出
摘要 目的:研究Raypex(5、Root SW两种电子根尖定位仪(electronic apexlocators,EAL)测量准确性。方法:采用琼脂凝胶和单根管离体牙构建EAL体外评价模型,使用两种EAL在同一种电介质中进行根管长度测量,记录测量值(measured length,ML)及测量值和实际值(actual length,AL)之间的差值(IF值)。结果采用Friedman分析。结果:IF值在±0.5mm范围内,Raypex(5为92.5%,Root SW为61.325%;IF值在±1.0mm范围内,Raypex(5为100%,Root SW为87.5%。在本实验中Raypex(5和Root SW测量准确性有显著性差异(P<0.001)。结论:电子根尖定位仪Raypex(5在本实验条件下可以较为准确测量实验牙根管长度。 Objective: To investigate the accuracy of two electronic apex locators(EAL). Methods: We use agar and single -canal extracted tooth to mimic the situation in vivo. Measured length and the distance between measured and actual length ( IF value) of two kinds of electronic :apex locator under one measurement environment were recorded. The Friedman test was applied to analyze the result. Results: The distance within limits of ±0.5mm was 92.5% for Raypex 5 and 61. 25% for Root SW. The distance within limits of ± 1.0mm was 100% for Raypex 5 and 87.5% for Root SW. There was a significant difference between two EALs regarding the accuracy of measurements. (P 〈 0.001 ). Conclusion: Electronic apex locator Raypex 5 can measure root canal length more accurately in this experimental environment.
出处 《口腔医学研究》 CAS CSCD 2007年第2期170-172,共3页 Journal of Oral Science Research
基金 湖北省青年杰出人才基金(编号:304161200)
关键词 电子根尖定位仪 准确性 根管 Electronic apex locator Accuracy Root canal
  • 相关文献

参考文献12

  • 1Kaufman AY,Keila S,Yoshpe M.Accuracy of a new apex locator:an in vitro study[J].Int Endod J,2002,35(2):186-192
  • 2Tamarut T,Kovacevic M,Uhac I.Detection of a transitional ion concentration zone during electronic measurement of root canal length:a study in vitro[J].Int Endod J,2000,33(4):374-380
  • 3Gordon MPJ,Chandler NP.Electronic apex locators[J].Int Endod J,2004,37(7):425-437
  • 4Pratten DH,McDonald NJ.Comparison of radiographic and electronic working lengths[J].J Endod,1996,22(4):173-176
  • 5Kim E,Lee SJ.Electronic apex locator[J].Dent Clin North Am,2004,48(1):35-54
  • 6Fan W,Fan B,Gutmann JL,et al.Evaluation of the accuracy of three electronic apex locators using glass tubules[J].Int Endod J,2006,39(2):127-135
  • 7Ebrahim AK,Yoshioka T,Kobayashi C,et al.The effects of file size,sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals:an in vitro study[J].Aust Dent J,2006,51(2):153-157
  • 8Green D.Stereomicroscopic study of 700 root apices of maxillary and mandibular posterior teeth[J].Oral Surg Oral Med Oral Pathol,1960,13:728-733
  • 9Dummer PM,McGinn JH,Rees DG.The position and topography of the apical canal constriction and apical foramen[J].Int Endod J,1984,17(4):192-198
  • 10Plotino G,Grande NM,Brigante L,et al.Ex vivo accuracy of three electronic apex locators:Root ZX,Elements Diagnostic Unit and Apex Locator and ProPex[J].Int Endod J,2006,39(5):408-414

同被引文献106

引证文献14

二级引证文献41

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部