导出
摘要 利用电化学修饰GOD工作电极及Ag/AgCl参考电极组装成了检测葡萄糖的GOD传感器.在0.7V的极化电位下,孩传感器对葡萄糖的响应电流小于1μA时,此电流与葡萄糖浓度成正比.传感器线性范围2.0×10-5至1.3×10-3mol/L,响应时间10s,稳定工作寿命150次,测量变异系数为2.73%。 Using a chemically modified GOD working electrode and the modified Pt electrode and a Ag/AgCl electrode, the sensor has been made for detectingglucose. Under positive potcntial of 0.7 V vs. Ag/AgCl, the oxidation currnt ofH202 is less than 1 μA, and it is proportional to the concentration of glucose inPBS. The GOD sensor's linear response range -2.0 × 10-3to 1 .3 ×10-3 mol/L, meanresponse time - 10 s, stable working life - 150 times and coefficient of variation(CV)- 2.73 %.
出处 《传感器技术》 CSCD 1997年第1期25-29,共5页 Journal of Transducer Technology
关键词 化学修饰 葡萄糖 氧化酶 传感器 Chemically modified electrode GOD Glucose sensor
  • 相关文献

参考文献5

  • 1黄文涛,酶应用手册,1989年,448页
  • 2李丽霞,生物工程学报,1987年,3卷,1期,46页
  • 3徐涛,临床检验基础,1986年,146页
  • 4Yao T,Anal Chim Acta,1983年,148期,27页
  • 5缪辉南,酶法分析手册,1983年,93,244页

同被引文献16

  • 1Guilbault GG,著.陈石根,译.酶法分析手册[M].上海:上海科学技术出版社,1983,98-103.
  • 2Updike SJ,Hioks GP,Nature[J].1967,214:986-989.
  • 3Yamaguchi.M,Mitsumori.M.,Kano.Y.Noninvasively Measuring Blood Glucose Using Saliva[J].Engineering in Medicine and Biology Magazine,IEEE,1998,17 (3):59-63.
  • 4Lynch SM,Bequette.B.W.Estimation based model predictive control of blood glucose in type I diabetics:a simulation study[C].Bio.engineering Conference,2001.Proceedings of the IEEE 27th Annual Northeast,31 March-1,2001.1767-1770.
  • 5Savage MB,Kun S,Harjunmaa H,et al.Development of a noninvasive blood glucose monitor:application of artificial neural networks for signal processing[J].American Control Conference,2002.Proceedings of the 2002,2002,5:4039-4043.
  • 6Guilbault G G, CHEN Shi-gen. Enzyme Analysis Manual [M].上海科学技术出版社,1983.
  • 7Vidal J C, Gareia-Ruiz E, Castillo J R. Design of a Multilayer Cholesterol Amperometrie Biosensor for Preparation and Use in Flow [J]. Systems Electroanalysis. 2001, 13(3): 229-235.
  • 8Vidal J C, Gareia-Ruiz E, Castillo J R. Strategies for the Improvement of an Amperometric Cholesterol Biosensor Based on Electropolymerization in Flow Systems: Use of Charge-Transfer Mediators and Platinization of the Electrode [J]. Journal of Pharmaceutical and Biomedical Analysis. 2000, 24(1): 51- 63.
  • 9Arun K R, Malhotra B D, Grover S K. Co-Immobilization of Cholesterol Oxidase and Horse Radish Peroxidasae in a Sol-Gel Film [J]. Analytiea Chimiea Aeta. 2000, 414(1): 43-50.
  • 10de Mattos I L, Lukaehoea L V, Gorton L, et al. Evaluationof Glucose Biosensors Based o Prussian Blue and Lyophilised,Crystalline Andeross-Linked Glucose Oxidase [J]. Talanta.2001, 54(5): 963-974.

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部