摘要
目的对自行研制的25#和27#四瓣叶、无支架、带腱索人工二尖瓣-超微孔膨体聚四氟乙烯人工心脏瓣膜(UPMV)进行体外脉动流检测,了解其体外脉动流力学特征。方法连续制作的25#UPMV和27#UPMV各6例使用TH-1200型人工心脏瓣膜体外脉动流试验台检测,循环率为70cycle/min。结果在流量为2、3、4、5、6L/min时,25#UPMV在体外脉动流下的平均跨瓣压差分别为(2.488±0.378)、(4.427±0.240)、(5.460±0.449)、(6.776±0.391)、(8.327±0.490)mmHg;有效瓣口面积分别为(1.430±0.333)、(1.993±0.208)、(2.260±0.477)、(3.204±0.174)、(3.652±0.158)cm2;返流百分比分别为(5.731±0.643)、(5.431±0.312)、(5.059±0.708)、(3.545±0.097)、(2.615±0.125)%。27#UPMV在体外脉动流下的平均跨瓣压差分别为(1.618±0.497)、(3.448±0.440)、(4.825±0.434)、(5.494±0.446)、(7.482±0.455)mmHg;有效瓣口面积分别为(1.773±0.364)、(2.113±0.305)、(2.409±0.295)、(3.326±0.417)、(4.522±0.445)cm2;返流百分比分别为(5.357±0.509)、(5.407±0.110)、(4.999±0.182)、(4.010±0.254)、(2.584±0.114)%。结论25#和27#UPMV在体外脉动流下的平均跨瓣压差、有效开口面积和返流百分比均符合国家标准。
Objective To test 25# and 27# ultramicroporous expanded polytetrafluo-roethylene mitral valve (UPMV) under pulsatile flow condition in vitro. Methods Six 25# and six 27#UPMV were tested using TH-1200 I-IV Prosthesis Pulsatile Flow Tester in vitro at the cycling rate of 70 cycle/min, with the systolic pressure maintained at about 16 kPa (120 mmHg), diastolic pressure at 10.7 kPa (80 mmHg), and the percentage of mean forward flow at 35%. The stimulant cardiac output was maintained at 2, 3, 4, 5 and 6 L/min, respectively, for the testing. Results The mean pressure difference for the 25#UPMV under stimulant cardiac output of 2, 3, 4, 5 and 6 L/min was 2.488±0.378, 4.427±0.240, 5.460±0.449, 6.776±0.391 and 8.327± 0.490 mmHg, and its effective valvular orifice was 1.430±0.333, 1.993±0.208, 2.260±0.477, 3.204±0.174 and 3.652±0.158 cm^2, respectively. The regurgitant fraction of the 25#UPMV under each stimulant cardiac output was (5.731 ±0.643)%, (5.431 ± 0.312)%, (5.059 ±0.708)%, (3.545 ±0.097)% and (2.615 ±0.125)%, respectively. The mean pressure difference of the 27# UPMV under each stimulant cardiac output was 1.618±0.497, 3.448±0.440, 4.825±0.434, 5.494±0.446 and 7.482±0.455 mmHg, effective valvularorificewas 1.773±0.364, 2.113±0.305, 2.409±0.295, 3.326±0.417and4.522±0.445cm2, and regurgitant fi-action was (5.357±0.509)%, (5.407 ±0.110)%, (4.999±0.182)0A, (4.010±0.254)% and (2.584±0.114)%, respectively. Conclusion The mean pressure difference, effective valvular orifice and regurgitant fraction of the UPMVs can measure up to the national criteria for artificial heart valve prosthesis of China.
出处
《南方医科大学学报》
CAS
CSCD
北大核心
2007年第5期660-662,共3页
Journal of Southern Medical University
基金
广东省科技计划项目(2002B11703)~~
关键词
人工心脏瓣膜
二尖瓣
超微孔膨体聚四氟乙烯人工二尖瓣
脉动流
artificial heart valve prosthesis
mitral valve
ultramicroporous expanded polytetrafluoroethylene mitral valve
pulsatile flow testing