期刊文献+

角度提取法双向定量血流速度仪的研制

ANGLE EXTRACTION TECHNIQUE FOR DIRECTIONAL AND QUANTITATIVE BLOOD FLOW VELOCITY MEASUREMENT WITH DOPPLER ULTRASOUND
原文传递
导出
摘要 提出了一种实现血流速度定量、双向检测的简单而且准确的方法。采用直接分离法实现血流速度的双向输出,利用角度提取法原理实现血流速度的定量输出。当偏角在-18~20°范围内变化时,此仪器的定量输出误差小于5%,其方向分离比大于42 dB。文中还给出了正常与异常的血流速度测试结果。 A simple but accurate method is developed to realized directional and quantita-tive measurement of blood flow velocity. Sideband filters are employed to get hi-direction output of blood flow. Angle extraction technique is suggested to obtainquantitative measurement. It shows that relative error is less than 5% while theangle deviation varies from -18 to +20°and separation of forward and reversevelocity is more than 42 dB. The blood flow velocity of normal and abnormal casesare also reported.
出处 《复旦学报(自然科学版)》 CAS CSCD 北大核心 1990年第4期442-447,共6页 Journal of Fudan University:Natural Science
关键词 血流速度仪 定理检测 角度提取法 blood flow, velocity measurement quantitative measurement, blood flow velocity, angle extraction.
  • 相关文献

参考文献2

  • 1邵谦明,1988年
  • 2王威琪,医疗器械,1985年,9卷,6期,1页

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...